Skip to content

epoc Blood Analysis BGEM Test Cards for use with the… recallZ-2034-2012 · Epocal Inc · initiated Jun 28, 2011

Class II Terminated All recalls posted in July 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Epocal Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.

FDA root cause category: Device Design

What the company is doing

Epocal Inc. sent a Voluntary Product Recall letter dated June 27, 2011 to the affected consignee. The letter identified the affected products, problem and actions to be taken. The letter ask that all products affected by this recall be sent back to Epocal and a replacement may be obtained. For questions call 1-613-738-6912.

Recall details

Recall number
Z-2034-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 28, 2011
Posted by the FDA
Jul 19, 2012
Recalling firm
Epocal Inc
Quantity
8,650
Product code
CGA
FDA submissions
K090109
Distribution
US nationwide Distribution in the state of CA.

Product description

epoc Blood Analysis BGEM Test Cards for use with the epoc(R) Blood Analysis System Product Usage: Usage: epoc Blood Analysis BGEM Test Cards are used with the epoc(R) Blood Analysis System which is a POC testing system

Code information (models, lots, serial numbers)

Catalog #CT-1004-00-00, lot number 07-11087-00

Other devices under the same FDA submission

1 device cleared under K090109 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page