Elecsys Troponin I Immunoassay recallZ-1387-2012 · Roche Diagnostics Operations, Inc. · initiated Mar 12, 2012
Terminated All recalls posted in April 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return the attached fax form via fax to 1-888-912-8457 and file this letter for future references. If you have any questions about the information contained in this letter, please contact Roche Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-428-2336.
Recall details
- Recall number
- Z-1387-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 12, 2012
- Posted by the FDA
- Apr 11, 2012
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 739 kits
- Product code
- MMI
- FDA submissions
- K082699
- Distribution
- Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, Algeria, Poland, Egypt, Qatar, Spain, Tunisia, France, Turkey, United Kingdom, Taiwan, India, Ukraine, Iran, Venezuela, Italy, South Africa, Jordan, and Armenia.
Product description
Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.
Code information (models, lots, serial numbers)
Elecsys Troponin I Immunoassay, 05094810160- lot 163176
Other devices under the same FDA submission
3 devices cleared under K082699 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Elecsys Troponin I STAT Immunoassay 5094798160Roche Diagnostics GmbH · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall3 reports
- Elecsys Troponin I Immunoassay 5094810160Roche Diagnostics GmbH · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall2 reports
- Elecsys Troponin T STAT Assay 4660307190Roche Diagnostics GmbH · IMMUNOASSAY METHOD, TROPONIN SUBUNIT6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 61249).
- Z-1388-2012 · TerminatedElecsys Troponin I STAT Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagn…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.