Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue… recallZ-1823-2012 · Arrow International Inc · initiated Jun 7, 2012

Class II Terminated All recalls posted in June 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Arrow is recalling the product due to the possibility of voids in the catheter hub which could lead to an interlumen crossover.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Arrow International sent a Urgent Medical Device Recall Notification letter dated June 11, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to take the following action: 1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog and lot numbers listed. 2. To return product complete the enclosed Recall Acknowledgement Form and fax to 1-800-343-2935, Attn: Customer Service. This will allow us to document the amount of the product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization ( RGA) Number and will provide instructions for the return of product to Arrow International. 3. If you have no affected stock, please complete t he enclosed Recall Acknowledgment Form and fax it to 1-800-343-2935, Attn: Customer Service. This will allow us to document your receipt of this letter. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.

Recall details

Recall number
Z-1823-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 7, 2012
Posted by the FDA
Jun 19, 2012
Recalling firm
Arrow International Inc
Quantity
1,085
Product code
FOZ
FDA submissions
K071538
Distribution
USA (nationwide) AL, AZ, AR, CA, CO, FL, GA, IL, IN, KS, LA, MA, MI, MO, NV, NH, NC, OH, PA, TX, VA, WA, and WI.

Product description

Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter - Catalog number CDC-42703-XP1A The multiple-lumen pressure injectable catheter permits venous access to central circulation and injection of contrast media. The ARROWg+ard(R) antimicrobial surface is intended to provide protection against catheter-related infections.

Code information (models, lots, serial numbers)

Lot number RF2033361

Other devices under the same FDA submission

231 devices cleared under K071538 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.