Skip to content

ARROW IPN915630Recalls & reported problems

Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Plus(R) Four-Lumen CVC · Class 2

All models of ARROW (1488)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10801902162158), received between Oct 31, 2023 and Aug 17, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

53 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 21, 2026Class IOpen, Classified

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    Z-2376-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Apr 21, 2026Class IOpen, Classified

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    Z-2384-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Material Frayed1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Unraveled Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50288697 ↗
  • MalfunctionOct 31, 2023Fluid/Blood Leak; Material Frayed; Fluid/Blood Leak; Material Frayed

    IT WAS REPORTED THAT: WHEN PRESSURE INJECTING CONTRACT THE CATHETER EXTENSION LINE STARTED LEAKING. ADDITIONAL INFORMATION: CVC WAS PLACED IN THE LEFT SUBCLAVIAN ON (B)(6) AND THE INCIDENT WAS ON THE (B)(6). AFTER THE ATTEMPT TO INJECT THE CONTRAST, THE AREA APPEARED TO BE FRAYED BETWEEN THE INSERTION SITE AND THE HUB/ADAPTER. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18037717 ↗

FDA clearance and classification

Cleared through 510(k) K071538 on Aug 30, 2007 (decision: Substantially Equivalent). Submitted by Arrow International, Inc. as “Arrow, Arrowg+Ard And Arrowg+Ard Blue Plus Pressure Injectable Central Venous Catheter”. The FDA review took 86 days. FDA record ↗

The FDA regulates product code FOZ (catheter, intravascular, therapeutic, short-term less than 30 days) under 21 CFR 880.5200 as a Class II device. 21 CFR 880.5200 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.

  • Z-1914-2026 · Mar 11, 2026 · Open, Classified
    ARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
  • Z-2123-2026 · Jan 7, 2026 · Open, Classified
    Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
  • Z-0003-2024 · Aug 10, 2023 · Open, Classified
    Arrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0002-2024 · Aug 10, 2023 · Open, Classified
    Arrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0352-2023 · Oct 27, 2022 · Terminated
    Pressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
  • Z-0351-2023 · Oct 27, 2022 · Terminated
    Arrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
  • Z-1604-2022 · May 25, 2022 · Open, Classified
    TELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
  • Z-1992-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Two-Lumen CVC Kit; Product Code: CDC-42802-P1A
  • Z-1989-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Quad-Lumen CVC Kit; Product Codes: CDC-42854-P1A and CDC-45854-P1A
  • Z-1979-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-15703-LD
  • Z-1993-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable ARROWg+ard Blue PLUS(R) Two-Lumen Central Venous Catheterization Kit; Product Code: ASK-42802-PWHC1
  • Z-1994-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Multi-Lumen CVC Kit; Product Codes: CDC-45703-P1A; CDC-45703-PB1A; CDC-42703-P1A; and CDC-15703-P1A
  • Z-1996-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; P
  • Z-1998-2018 · Apr 11, 2018 · Terminated
    Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-427
  • Z-1981-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Multi-Lumen Central Venous Catheterization Kit; Product Code: ASK-15703-PVH and ASK-15703-PNYP

…and 38 more.

Common questions about the ARROW IPN915630

Has the ARROW IPN915630 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2376-2026 (Class I), began on Apr 21, 2026: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ARROW IPN915630?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Oct 31, 2023 and Aug 17, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (1), “Material Frayed” (1) and “Unraveled Material” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ARROW IPN915630?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ARROW IPN915630 FDA approved?

It was cleared through 510(k) K071538 on Aug 30, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ARROW IPN915630, and is it still sold?

The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN915630). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page