I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ recallZ-2133-2012 · Abbott Point Of Care Inc. · initiated Jan 1, 2011

Class II Terminated All recalls posted in August 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Abbott Point of Care, Inc sent a Urgent Recall Notice letter dated January 14, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of Cartridges from the identified lot. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card. If you have any questions regarding this information please contact your Abbott Point of Care Technical Support at 800-366-8020 Option 1, or your Abbott Point of Care representative.

Recall details

Recall number
Z-2133-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 1, 2011
Posted by the FDA
Aug 2, 2012
Recalling firm
Abbott Point Of Care Inc.
Quantity
52,750 cartridges
Product code
CHL
FDA submissions
K940918
Distribution
USA ( nationwide )

Product description

I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.

Code information (models, lots, serial numbers)

List number 03P76-25 Lot number S10246A

Other devices under the same FDA submission

5 devices cleared under K940918 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.