Siemens ADVIA Centaur Multi-Diluent 1 - Used for automated… recallZ-2313-2012 · Siemens Healthcare Diagnostics, Inc · initiated Jul 25, 2012

Class II Terminated All recalls posted in September 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Decrease in onboard dilution recovery when using Multi-Diluent 1 that has been stored on board the ADVIA Centaur and ADVIA Centaur XP systems for for BNP and TSH3 Ultra assays

FDA root cause category: Nonconforming Material/Component

What the company is doing

Siemens sent an "URGENT DEVICE RECALL NOTICE" dated July 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Field Correction Effectiveness Check Response Form was included for customers to complete and return via fax to (302) 631-7597. Contact your local Sales Representative for questions regarding this notice.

Recall details

Recall number
Z-2313-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 25, 2012
Posted by the FDA
Sep 6, 2012
Recalling firm
Siemens Healthcare Diagnostics, Inc
Quantity
2 Packs = 15879 and 6 Packs = 7414
Product code
NBC
FDA submissions
K031038, K083844
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Argentina, Australia, Brazil, Chile, China, Croatia, Denmark, Ecuduar, Egypt,Guatemala, Hong Kong,India,Ireland, Israel,India, Japan,Korea, Mexico, Malyasia, Peru, Pakistan, Saudi Arabia,Singapore, Thailand,Uruguay, Viet Nam, and South Africa.

Product description

Siemens ADVIA Centaur Multi-Diluent 1 - Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems. Catalog Numbers: 2 Pack REF 07907174 (110312) 6 Pack REF 07293184 (110313). Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Lot # Expiry Date 2 Pack 76129 22 JUN 2012  77542 22 JUN 2012  78696 22 JUN 2012  80828 26 SEP 2012  80857 26 SEP 2012  81441 26 SEP 2012  82189 26 SEP 2012  83647 26 SEP 2012  84772 26 SEP 2012  86037 13 JAN 2013  86783 13 JAN 2013  88712 13 JAN 2013  90056 13 JAN 2013  91002 13 JAN 2013  93126 13 JAN 2013  6 Pack 76102 22 JUN 2012  76432 22 JUN 2012  77075 22 JUN 2012  78264 22 JUN 2012  79037 22 JUN 2012  81006 26 SEP 2012  82539 26 SEP 2012  83321 26 SEP 2012  84123 26 SEP 2012  85795 13 JAN 2013  86698 13 JAN 2013  88318 13 JAN 2013  89286 13 JAN 2013  90490 13 JAN 2013  91672 13 JAN 2013  93129 13 JAN 2013

Other devices under the same FDA submission

2 devices cleared under K031038, K083844 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.