Qkc1692e recallZ-2284-2012 · Steris Corporation · initiated Aug 10, 2012

Class II Terminated All recalls posted in August 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

FDA root cause category: Device Design

What the company is doing

The firm, Steris, sent a "Voluntary Field Correction: Re-labeling of QKC1692E" letter on August 10, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the steps outlined in the letter to re-label the affected QKC1692E(s) and return the completed form by September 30. 2012 to: Regulatory Affairs Specialist via [email protected] or via fax at (440) 350-7082. If you have any questions, contact Director, Regulatory Affairs at 440-392-7519.

Recall details

Recall number
Z-2284-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 10, 2012
Posted by the FDA
Aug 28, 2012
Recalling firm
Steris Corporation
Quantity
64 units
Product code
MED
FDA submissions
K090036
Distribution
Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.

Product description

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Code information (models, lots, serial numbers)

Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190

Other devices under the same FDA submission

4 devices cleared under K090036 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.