QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator recallZ-0160-2013 · Maquet Cardiovascular Us Sales, Llc · initiated Jun 12, 2012
Class II Terminated All recalls posted in October 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Maquet sent a Field Safety Notices and Confirmation Forms dated June 13, 2012 to all affected customers. A Follow-up communication dated July 20, 2012 was mailed to all affected US customers. The Field Safety Notice identifies the problem, affected product, risk factors and actions to be taken by the customer. Customers were instructed to complete the Confirmation form and return in the self-addressed prepaid envelop. For questions call 201-995-8968. Maquet issued an updated Field Safety Notice/Field Safety Notice Confirmation Form dated 2/14/13 via Fed Ex on 2/18/2013. Maquet sent out an Updated Field Safety Notice/field Safety Notice Confirmation Form dated 4/4/2013 in order to notify customers they would be providing safety clamps.
Recall details
- Recall number
- Z-0160-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 12, 2012
- Posted by the FDA
- Oct 26, 2012
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 23,203 units
- Product code
- DTZ
- FDA submissions
- K082117, K090511, K090689, K093522, K103191, K112360
- Distribution
- US Nationwide Distribution
Product description
QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895. QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U. HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050. HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050. Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX. Note: an X was used to represent variations of catalogue numbers.
Other devices under the same FDA submission
4 devices cleared under K082117, K090511, K090689, K093522, K103191, K112360 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- NA HLS Set Advanced 7.0, Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page