QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator recallZ-0160-2013 · Maquet Cardiovascular Us Sales, Llc · initiated Jun 12, 2012

Class II Terminated All recalls posted in October 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Maquet sent a Field Safety Notices and Confirmation Forms dated June 13, 2012 to all affected customers. A Follow-up communication dated July 20, 2012 was mailed to all affected US customers. The Field Safety Notice identifies the problem, affected product, risk factors and actions to be taken by the customer. Customers were instructed to complete the Confirmation form and return in the self-addressed prepaid envelop. For questions call 201-995-8968. Maquet issued an updated Field Safety Notice/Field Safety Notice Confirmation Form dated 2/14/13 via Fed Ex on 2/18/2013. Maquet sent out an Updated Field Safety Notice/field Safety Notice Confirmation Form dated 4/4/2013 in order to notify customers they would be providing safety clamps.

Recall details

Recall number
Z-0160-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 12, 2012
Posted by the FDA
Oct 26, 2012
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Quantity
23,203 units
Product code
DTZ
FDA submissions
K082117, K090511, K090689, K093522, K103191, K112360
Distribution
US Nationwide Distribution

Product description

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895.   QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U.  HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050.  HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050.  Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX.  Note: an X was used to represent variations of catalogue numbers.

Other devices under the same FDA submission

4 devices cleared under K082117, K090511, K090689, K093522, K103191, K112360 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.