Skip to content

HLS Set BEQ-HLS 5050 USARecalls & reported problems

Maquet Cardiopulmonary AG · HLS Set Advanced 5.0, Small Adult, BIOLINE Coating · Class 2

All models of HLS Set (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04058863076355), received between Aug 13, 2024 and Jun 11, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

3 FDA recall records name this device's identifier, 3 still open.

9 other recalls cover devices cleared under the same FDA submission (K112360); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 16, 2024Class IIOpen, Classified

    The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.

    Z-1076-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Feb 13, 2023Class IIOpen, Classified

    Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.

    Z-1188-2023 · Root cause: Process control · Recall details · FDA record ↗

  3. Jul 22, 2022Class IIOpen, Classified

    The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

    Z-1766-2022 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Partial Blockage2
  • Device Contamination with Chemical or Other Material1
  • Leak/Splash1
  • Pressure Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2026Pressure Problem

    THE EVENT OCCURRED IN THE USA PRIOR TO USE ON (B)(6) 2026. IT WAS REPORTED THAT THE HLS SET HAD PRESSURE READING ISSUES AFTER SET-UP. P VEN COULD NOT BE READ, IT WAS DISPLAYED AS DASHES. THE AFFECTED HLS SET WAS PRIMED ON (B)(6) 2026. THE PRESSURES ZEROED WHILE THE SET WAS FREE OF LIQUIDS/PRIMING SOLUTION. NO VISIBLE DAMAGE WAS SEEN ON THE SENSOR HOUSING OR GREEN FLEXLINE. IN ORDER TO RECTIFY THE ISSUE, THE HLS CABLE WAS REPLACED. THE AFFECTED HLS SET WAS AS WELL CONNECTED TO A DIFFERENT CARDIOHELP DEVICE. HOWEVER, THE ISSUE COULD NOT BE SOLVED. THE CARDIOHELP DEVICES COULD BE EXCLUDED AS FAULT. THE AFFECTED HLS SET WAS NOT CONNECTED TO A PATIENT. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS A PRESSURE READING ISSUE CAN LEAD TO A PUMP STOP, IF THE INTERVENTION IS SET BY THE USER, A REPORT IS REQUIRED. COMPLAINT ID# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25467577 ↗
  • MalfunctionDec 9, 2025Partial Blockage; Fluid/Blood Leak; Partial Blockage; Fluid/Blood Leak

    COMPLAINT ID# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756333 ↗
  • MalfunctionOct 1, 2024Partial Blockage; Fluid/Blood Leak; Partial Blockage; Fluid/Blood Leak

    COMPLAINT ID# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20342572 ↗
  • MalfunctionAug 13, 2024Leak/Splash; Device Contamination with Chemical or Other Material; Leak/Splash; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE CUSTOMER IS TRIALING A NEW HEATER WITH THE ECMO DEVICES. THE NEW HEATERS (NOT A GETINGE PRODUCT) ARE CLEANED WITH VIREX PLUS. THE CHEMICAL IS RINSED OUT DURING THE CLEANING CYCLE, BUT ONE OF THE HEATERS APPEARED TO HAVE SOME VIREX PLUS STILL PRESENT. THIS WAS DISCOVERED ONCE PLACED ON A PATIENT, WHEN FOAM WAS DISCOVERED COMING OUT OF THE GAS OUTLET OF THE HLS MODULE. THERE IS NO HARM TO THE PATIENT. THE CUSTOMER IS PLANNING ON CHANGING THE HLS SET. AFTER THAT EVENT THE CUSTOMER IS ALSO PLANNING TO SWITCH TO ISOPROPYL ALCOHOL FOR THE CLEANING AGENT. NO HARM TO ANY PERSON HAS BEEN REPORTED. ON 2024-08-08 THE INFORMATION WAS RECEIVED THAT THE HLS SET WAS EXCHANGED 36 HOURS LATER AS EXCESSIVE FLUID CONTINUED TO DROP FROM THE GAS OUTLET EVEN AFTER THE FOAM CEASED. THE FOAM WAS CLEARED AFTER 30 MINUTES. THIS WAS MITIGATED BY CHANGING THE HEATER DEVICE. AS THE FAILURE OCCURRED DURING TREATMENT, AND THE DEVICE WAS EXCHANGED THE EVENT REPRESENTS A RISK FOR THE PATIENT, THEREFORE A REPORT IS REQUIRED. COMPLAINT ID# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19972392 ↗

FDA clearance and classification

Cleared through Special 510(k) K112360 on Sep 15, 2011 (decision: Substantially Equivalent). Submitted by Maquet Cardiopulmonary, AG as “Quadrox-I Microporporous Membrane Oxgenator Series And Qudrox-Id Diffusion Membrane Oxygenators Series”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K112360. One submission often covers many models, so check the product and lot details in each record.

  • Z-1077-2024 · Jan 16, 2024 · Open, Classified
    BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.
  • Z-1767-2022 · Jul 22, 2022 · Open, Classified
    BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70106.9078. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP Sy
  • Z-1764-2022 · Jul 22, 2022 · Open, Classified
    BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70105.2794. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP Sy
  • Z-1765-2022 · Jul 22, 2022 · Open, Classified
    BEQ-HLS 5050 USA; HLS Set Advanced 5.0 Part No. 70105.2797. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP Sy
  • Z-0633-2021 · Nov 20, 2020 · Terminated
    HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - Product Usage: These devices are used with the CARDIOHELP system in order to
  • Z-0632-2021 · Nov 20, 2020 · Terminated
    HLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 - Product Usage: These devices are used with the CARDIOHELP system in orde
  • Z-2149-2019 · Jul 11, 2019 · Terminated
    Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794
  • Z-2262-2013 · Aug 15, 2013 · Terminated
    QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracor
  • Z-0160-2013 · Jun 12, 2012 · Terminated
    QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended

Common questions about the HLS Set BEQ-HLS 5050 USA

Has the HLS Set BEQ-HLS 5050 USA been recalled?

Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-1076-2024 (Class II), began on Jan 16, 2024: The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the HLS Set BEQ-HLS 5050 USA?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Aug 13, 2024 and Jun 11, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (2), “Partial Blockage” (2) and “Device Contamination with Chemical or Other Material” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HLS Set BEQ-HLS 5050 USA?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HLS Set BEQ-HLS 5050 USA FDA approved?

It was cleared through Special 510(k) K112360 on Sep 15, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HLS Set BEQ-HLS 5050 USA, and is it still sold?

The FDA's device database lists Maquet Cardiopulmonary AG as the company responsible for it (brand HLS Set, model BEQ-HLS 5050 USA). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Oxygenator, Cardiopulmonary Bypass devices

All Oxygenator, Cardiopulmonary Bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page