Hologic Fluoroscan Mini C-Arm Model with InSight2, Software… recallZ-2452-2012 · Hologic, Inc. · initiated Sep 14, 2012
Class II Terminated All recalls posted in September 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180 around vertical axis as well as 180 around a horizontal axis.
FDA root cause category: Software design
What the company is doing
Hologic issued an Urgent: Medical Device recall letter dated September 14, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Halogen corrected the anomaly and they provided the customer with the revised version of the software. Customers were asked to update their systems in accordance with the instructions provided with the software included with the letter. Customers were also asked to complete the attached acknowledgement form and fax or mail it to the firm at their earliest convenience. For any questions customers were instructed to call Hologic's Help Desk at 1-800-321-4659. For questions regarding this recall call 781-999-7300.
Recall details
- Recall number
- Z-2452-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 14, 2012
- Posted by the FDA
- Sep 25, 2012
- Recalling firm
- Hologic, Inc.
- Quantity
- 414 units
- Product code
- JAA
- FDA submissions
- K051025
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.
Product description
Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.
Code information (models, lots, serial numbers)
Software version 4.0 and 4.0.1
Other devices under the same FDA submission
1 device cleared under K051025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- InSight 2Hologic, Inc. · System, x-ray, fluoroscopic, image-intensified1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 63180).
- Z-2453-2012 · TerminatedHologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with ge…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page