Hologic Fluoroscan Mini C-Arm Model with InSight- FD… recallZ-2453-2012 · Hologic, Inc. · initiated Sep 14, 2012

Class II Terminated All recalls posted in September 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180 around vertical axis as well as 180 around a horizontal axis.

FDA root cause category: Software design

What the company is doing

Hologic issued an Urgent: Medical Device recall letter dated September 14, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Halogen corrected the anomaly and they provided the customer with the revised version of the software. Customers were asked to update their systems in accordance with the instructions provided with the software included with the letter. Customers were also asked to complete the attached acknowledgement form and fax or mail it to the firm at their earliest convenience. For any questions customers were instructed to call Hologic's Help Desk at 1-800-321-4659. For questions regarding this recall call 781-999-7300.

Recall details

Recall number
Z-2453-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 14, 2012
Posted by the FDA
Sep 25, 2012
Recalling firm
Hologic, Inc.
Quantity
46 units
Product code
JAA
FDA submissions
K051025
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.

Product description

Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

Code information (models, lots, serial numbers)

Software version 5.0.1and 5.0.2

Other devices under the same FDA submission

1 device cleared under K051025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 63180).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page