STA Satellite Automated Multi-Parameter Analyzer recallZ-0252-2013 · Diagnostica Stago, Inc. · initiated Jun 15, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite.

FDA root cause category: Software design

What the company is doing

Diagnostica Stago sent an Urgent Field Safety Notice dated June 12, 2012, to all affected customers via email return receipt requested on June 15, 2012. Customers were instructed to follow the instructions in the notification and complete the Customer Verification form and return to the Hotline by fax to 973-644-9348. For additional information, customers were instructed to contact the Hotline at 800-725-0607. For questions regarding this recall call 973-631-1200, ext 2044.

Recall details

Recall number
Z-0252-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 15, 2012
Posted by the FDA
Nov 6, 2012
Recalling firm
Diagnostica Stago, Inc.
Quantity
232 units US; 5 units Canada
Product code
JPA
FDA submissions
K082248
Distribution
Worldwide Distribution - USA (nationwide) and Canada

Product description

STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Code information (models, lots, serial numbers)

K082248 D089976 Ref 58104 units

Other devices under the same FDA submission

1 device cleared under K082248 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.