IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part… recallZ-0428-2013 · Innovative Magnetic Resonance Imaging Systems Inc. · initiated Oct 19, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Innovative Magnetic Resonance Imaging Systems Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.

FDA root cause category: Labeling design

What the company is doing

Consignees were sent on 10/19/12 a IMRIS "Urgent Medical Device Correction" letter dated October 16. 2012. The letter described the product and the problem involved in the recall. It also provided changes to the Operator's Manual and labeling which addresses the issue. Customers may contact 1-866-475-0525.

Recall details

Recall number
Z-0428-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 19, 2012
Posted by the FDA
Nov 21, 2012
Recalling firm
Innovative Magnetic Resonance Imaging Systems Inc.
Quantity
23
Product code
LNH
FDA submissions
K061916, K071099, K083137
Distribution
Nationwide Distribution, including the states of FL, GA, IN, MD,MA, MI, MN, MO, TN, UT, and VA.

Product description

IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.

Code information (models, lots, serial numbers)

10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513

Other devices under the same FDA submission

3 devices cleared under K061916, K071099, K083137 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.