Skip to content

Neuro III-SV iMRI System (UDI 00857534006202)Recalls & reported problems

IMRIS Imaging, Inc. · Neuro III-SV Multi-Room Imaging Suite MR30LS Enhancement OR-DR, 116189-000 · Class 2

All models of Neuro III-SV iMRI System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K083137); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 30, 2025Class IOpen, Classified

    There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    Z-0161-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083137. One submission often covers many models, so check the product and lot details in each record.

  • Z-0163-2026 · Sep 30, 2025 · Open, Classified
    IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
  • Z-0162-2026 · Sep 30, 2025 · Open, Classified
    IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
  • Z-0164-2026 · Sep 30, 2025 · Open, Classified
    IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
  • Z-1098-2016 · Feb 10, 2016 · Terminated
    Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).
  • Z-0428-2013 · Oct 19, 2012 · Terminated
    IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.

Other System, Nuclear Magnetic Resonance Imaging devices

All System, Nuclear Magnetic Resonance Imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page