qube Compact Monitor (Patient Monitor) 91390. The monitor… recallZ-0403-2013 · Spacelabs Healthcare, Llc · initiated Oct 19, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated October 19, 2012. The letter identified the product, problem, and actions to be taken by the customers. Contact Spacelabs Healthcare at 1-800522-7025 option 2 for questions regarding this notice.

Recall details

Recall number
Z-0403-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 19, 2012
Posted by the FDA
Nov 20, 2012
Recalling firm
Spacelabs Healthcare, Llc
Quantity
742 total devices distributed; 530 with the U.S. and 212 Internationally.
Product code
MHX
FDA submissions
K120616
Distribution
Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as replacements to a French customer whose devices had been returned to support the firm's investigation. The customer had already received the Urgent Medical Device correction, customer letter and required Spacelabs to provide these devices with the understanding that they would require the field corrective action

Product description

qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.1 inches) LCD display with 1024 x 768 resolution. The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by 2 Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECU, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed. The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow network-based applications to open windows and display information on other networked monitors. The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Code information (models, lots, serial numbers)

Show the full code information (10k characters)

Serial numbers for devices distributed in the US:  1390-000102, 1390-000103, 1390-000104, 1390-000105, 1390-000106, 1390-000107, 1390-000108, 1390-000109, 1390-000110, 1390-000111, 1390-000112, 1390-000113, 1390-000114, 1390-000115, 1390-000116, 1390-000119, 1390-000120, 1390-000150, 1390-000151, 1390-000152, 1390-000153, 1390-000154, 1390-000155, 1390-000156, 1390-000157, 1390-000158, 1390-000159, 1390-000160, 1390-000161, 1390-000162, 1390-000163, 1390-000164, 1390-000165, 1390-000166, 1390-000167, 1390-000168, 1390-000169, 1390-000170, 1390-000171, 1390-000172, 1390-000173, 1390-000174, 1390-000175, 1390-000176, 1390-000177, 1390-000178, 1390-000179, 1390-000180, 1390-000181, 1390-000182, 1390-000183, 1390-000184, 1390-000185, 1390-000186, 1390-000187, 1390-000188, 1390-000189, 1390-000190, 1390-000191, 1390-000192, 1390-000193, 1390-000195, 1390-000196, 1390-000197, 1390-000198, 1390-000199, 1390-000200, 1390-000201, 1390-000202, 1390-000203, 1390-000204, 1390-000205, 1390-000206, 1390-000207, 1390-000208, 1390-000209, 1390-000210, 1390-000211, 1390-000212, 1390-000213, 1390-000214, 1390-000215, 1390-000216, 1390-000217, 1390-000218, 1390-000219, 1390-000220, 1390-000221, 1390-000222, 1390-000223, 1390-000224, 1390-000225, 1390-000226, 1390-000227, 1390-000228, 1390-000229, 1390-000230, 1390-000231, 1390-000232, 1390-000233, 1390-000234, 1390-000235, 1390-000236, 1390-000237, 1390-000238, 1390-000239, 1390-000240, 1390-000241, 1390-000242, 1390-000243, 1390-000244, 1390-000245, 1390-000246, 1390-000247, 1390-000248, 1390-000249, 1390-000250, 1390-000251, 1390-000252, 1390-000253, 1390-000254, 1390-000255, 1390-000256, 1390-000257, 1390-000258, 1390-000259, 1390-000260, 1390-000261, 1390-000262, 1390-000263, 1390-000264, 1390-000266, 1390-000267, 1390-000268, 1390-000269, 1390-000270, 1390-000271, 1390-000272, 1390-000273, 1390-000274, 1390-000275, 1390-000276, 1390-000277, 1390-000278, 1390-000279, 1390-000280, 1390-000281, 1390-000282, 1390-000283, 1390-000284, 1390-000285, 1390-000286, 1390-000287, 1390-000288, 1390-000289, 1390-000296, 1390-000297, 1390-000298, 1390-000299, 1390-000300, 1390-000301, 1390-000302, 1390-000304, 1390-000305, 1390-000307, 1390-000308, 1390-000310, 1390-000311, 1390-000312, 1390-000314, 1390-000315, 1390-000316, 1390-000317, 1390-000318, 1390-000319, 1390-000321, 1390-000322, 1390-000323, 1390-000325, 1390-000341, 1390-000342, 1390-000343, 1390-000344, 1390-000345, 1390-000346, 1390-000347, 1390-000348, 1390-000349, 1390-000350, 1390-000351, 1390-000352, 1390-000353, 1390-000354, 1390-000355, 1390-000356, 1390-000357, 1390-000358, 1390-000359, 1390-000360, 1390-000361, 1390-000362, 1390-000363, 1390-000364, 1390-000365, 1390-000366, 1390-000367, 1390-000368, 1390-000369, 1390-000370, 1390-000371, 1390-000372, 1390-000373, 1390-000374, 1390-000375, 1390-000376, 1390-000377, 1390-000378, 1390-000379, 1390-000380, 1390-000381, 1390-000382, 1390-000383, 1390-000384, 1390-000385, 1390-000386, 1390-000387, 1390-000388, 1390-000389, 1390-000390, 1390-000391, 1390-000392, 1390-000393, 1390-000394, 1390-000420, 1390-000421, 1390-000422, 1390-000423, 1390-000424, 1390-000425, 1390-000426, 1390-000427, 1390-000428, 1390-000429, 1390-000430, 1390-000431, 1390-000432, 1390-000433, 1390-000434, 1390-000435, 1390-000436, 1390-000437, 1390-000438, 1390-000439, 1390-000440, 1390-000441, 1390-000442, 1390-000443, 1390-000444, 1390-000445, 1390-000446, 1390-000447, 1390-000448, 1390-000449, 1390-000450, 1390-000451, 1390-000452, 1390-000453, 1390-000454, 1390-000455, 1390-000456, 1390-000457, 1390-000458, 1390-000459, 1390-000460, 1390-000461, 1390-000462, 1390-000463, 1390-000464, 1390-000465, 1390-000467, 1390-000468, 1390-000469, 1390-000470, 1390-000483, 1390-000564, 1390-000582, 1390-000583, 1390-000584, 1390-000585, 1390-000586, 1390-000587, 1390-000588, 1390-000589, 1390-000590, 1390-000591, 1390-000592, 1390-000593, 1390-000594, 1390-000595, 1390-000597, 1390-000598, 1390-000599, 1390-000600, 1390-000601, 1390-000602, 1390-000603, 1390-000606, 1390-000614, 1390-000615, 1390-000616, 1390-000623, 1390-000624, 1390-000625, 1390-000626, 1390-000627, 1390-000628, 1390-000629, 1390-000630, 1390-000631, 1390-000632, 1390-000633, 1390-000634, 1390-000635, 1390-000636, 1390-000637, 1390-000638, 1390-000639, 1390-000640, 1390-000641, 1390-000642, 1390-000643, 1390-000644, 1390-000645, 1390-000646, 1390-000647, 1390-000648, 1390-000649, 1390-000650, 1390-000651, 1390-000652, 1390-000653, 1390-000654, 1390-000655, 1390-000656, 1390-000657, 1390-000658, 1390-000659, 1390-000660, 1390-000661, 1390-000662, 1390-000663, 1390-000664, 1390-000665, 1390-000666, 1390-000667, 1390-000668, 1390-000669, 1390-000670, 1390-000671, 1390-000672, 1390-000673, 1390-000674, 1390-000675, 1390-000676, 1390-000677, 1390-000678, 1390-000679, 1390-000680, 1390-000681, 1390-000682, 1390-000683, 1390-000684, 1390-000685, 1390-000686, 1390-000687, 1390-000688, 1390-000698, 1390-000699, 1390-000700, 1390-000701, 1390-000702, 1390-000703, 1390-000704, 1390-000705, 1390-000706, 1390-000707, 1390-000708, 1390-000710, 1390-000712, 1390-000713, 1390-000714, 1390-000720, 1390-000721, 1390-000728, 1390-000729, 1390-000730, 1390-000731, 1390-000732, 1390-000733, 1390-000734, 1390-000735, 1390-000736, 1390-000737, 1390-000738, 1390-000742, 1390-000743, 1390-000744, 1390-000745, 1390-000746, 1390-000747, 1390-000748, 1390-000749, 1390-000750, 1390-000751, 1390-000752, 1390-000753, 1390-000754, 1390-000755, 1390-000756, 1390-000757, 1390-000758, 1390-000759, 1390-000760, 1390-000761, 1390-000762, 1390-000763, 1390-000764, 1390-000765, 1390-000766, 1390-000767, 1390-000768, 1390-000769, 1390-000775, 1390-000776, 1390-000777, 1390-000778, 1390-000779, 1390-000780, 1390-000781, 1390-000782, 1390-000792, 1390-000793, 1390-000794, 1390-000795, 1390-000796, 1390-000800, 1390-000801, 1390-000802, 1390-000803, 1390-000804, 1390-000805, 1390-000806, 1390-000807, 1390-000808, 1390-000809, 1390-000810, 1390-000811, 1390-000812, 1390-000813, 1390-000814, 1390-000815, 1390-000816, 1390-000817, 1390-000818, 1390-000819, 1390-000820, 1390-000821, 1390-000822, 1390-000823, 1390-000824, 1390-000825, 1390-000826, 1390-000827, 1390-000828, 1390-000829, 1390-000830, 1390-000831, 1390-000832, 1390-000833, 1390-000834, 1390-000835, 1390-000836, 1390-000837, 1390-000838, 1390-000839, 1390-000840, 1390-000841, 1390-000842, 1390-000843, 1390-000844, 1390-000845, 1390-000846, 1390-000847, 1390-000848, 1390-000849, 1390-000850, 1390-000851, 1390-000852, 1390-000853, 1390-000854, 1390-000855, 1390-000856, 1390-000857, 1390-000858, 1390-000859, 1390-000860, 1390-000861, 1390-000862, 1390-000863, 1390-000870, 1390-000871, 1390-000872, 1390-000887, 1390-000888, 1390-000889, 1390-000890, 1390-000891, 1390-000892, 1390-000893, 1390-000894, 1390-000895, 1390-000896, 1390-000897, 1390-000898, 1390-000899, 1390-000900, 1390-000901, 1390-000902, 1390-000903, and 1390-000904.   Serial numbers for devices distributed INTERNATIONAL: On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913.  1390-000290, 1390-000291, 1390-000337, 1390-000338, 1390-000339, 1390-000340, 1390-000607, 1390-000608, 1390-000609, 1390-000610, 1390-000689, 1390-000690, 1390-000691, 1390-000692, 1390-000693, 1390-000695, 1390-000536, 1390-000537, 1390-000709, 1390-000722, 1390-000723, 1390-000724, 1390-000725, 1390-000879, 1390-000880, 1390-000881, 1390-000905, 1390-000906, 1390-000907, 1390-000908, 1390-000909, 1390-000910 1390-000009, 1390-000010, 1390-000011, 1390-000012, 1390-000013, 1390-000014, 1390-000015, 1390-000016, 1390-000017, 1390-000018, 1390-000046, 1390-000048, 1390-000049, 1390-000051, 1390-000052, 1390-000053, 1390-000056, 1390-000057, 1390-000059, 1390-000060, 1390-000061, 1390-000062, 1390-000063, 1390-000101, 1390-000117, 1390-000118, 1390-000125, 1390-000127, 1390-000128, 1390-000138, 1390-000139, 1390-000140, 1390-000141, 1390-000142, 1390-000143, 1390-000144, 1390-000145, 1390-000146, 1390-000147, 1390-000148, 1390-000149, 1390-000326, 1390-000327, 1390-000328, 1390-000329, 1390-000330, 1390-000331, 1390-000332, 1390-000333, 1390-000334, 1390-000335, 1390-000336, 1390-000466, 1390-000535, 1390-000552, 1390-000568, 1390-000605, 1390-000611, 1390-000612, 1390-000613, 1390-000715, 1390-000716, 1390-000739, 1390-000740, 1390-000741, 1390-000770, 1390-000787, 1390-000788, 1390-000789, 1390-000790, 1390-000791, 1390-000797, 1390-000798, 1390-000876, 1390-000877, 1390-000878, 1390-1390-000047, 1390-00002, 1390-000618, 1390-000619, 1390-000620, 1390-000621, 1390-000622, 1390-000509, 1390-000510, 1390-000511, 1390-000512, 1390-000513, 1390-000514, 1390-000553, 1390-000554, 1390-000121, 1390-000524, 1390-000771, 1390-000772, 1390-000292, 1390-000293, 1390-000294, 1390-000295, 1390-000911, 1390-000555, 1390-000556, 1390-000557, 1390-000558, 1390-000559, 1390-000560, 1390-000561, 1390-000562, 1390-000563, 1390-000565, 1390-000566, 1390-000567, 1390-000569, 1390-000570, 1390-000571, 1390-000572, 1390-000573, 1390-000574, 1390-000575, 1390-000576, 1390-000577, 1390-000578, 1390-000579, 1390-000580, 1390-000581 1390-000395, 1390-000396, 1390-000397, 1390-000398, 1390-000399, 1390-000400, 1390-000401, 1390-000402, 1390-000403, 1390-000404, 1390-000405, 1390-000406, 1390-000407, 1390-000408, 1390-000409, 1390-000410, 1390-000411, 1390-000412, 1390-000413, 1390-000414, 1390-000415, 1390-000416, 1390-000417, 1390-000418, 1390-000419, 1390-000604, 1390-000130, 1390-000131, 1390-000132, 1390-000133, 1390-000135, 1390-000136, 1390-000137, 1390-000493, 1390-000496, 1390-000497, 1390-000498, 1390-000499, 1390-000500, 1390-000501, 1390-000502, 1390-000525, 1390-000526, 1390-000527, 1390-000528, 1390-000529, 1390-000530, 1390-000531, 1390-000532, 1390-000533, 1390-000534, 1390-000596, 1390-000617, 1390-000799, and 1390-000696.

Other devices under the same FDA submission

3 devices cleared under K120616 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.