Skip to content

Medfusion Model 3500 Syringe Pump, Model #3500-415 and… recallZ-0996-2013 · Smiths Medical ASD, Inc. · initiated May 14, 2010

Class II Terminated All recalls posted in March 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.

FDA root cause category: Device Design

What the company is doing

On May 14th, 2010, Smiths Medical contacted the facility that received the pumps via e-mail and requested that they immediately return the pumps. The pumps were returned per Returned Goods Authorization (RGA) and exchanged for replacement pumps manufactured with plunger rod seal G6000077.

Recall details

Recall number
Z-0996-2013
Recall class
Class II
Status
Terminated
Initiated by the company
May 14, 2010
Posted by the FDA
Mar 26, 2013
Recalling firm
Smiths Medical ASD, Inc.
Quantity
6
Product code
FRN
FDA submissions
K040899
Distribution
Distributed in the state of NH.

Product description

Medfusion Model 3500 Syringe Pump, Model #3500-415 and 3500-500. Intended for use in critical care areas for the administration of fluids requiring precisely controlled infusion rates.

Code information (models, lots, serial numbers)

Lot/Serial Numbers: M76369, M76370, M76373, M76374, M76375, and M76377.

Other devices under the same FDA submission

1 device cleared under K040899 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page