Medical device recalls: March 2013
The FDA posted 124 medical device recall records in March 2013, from 96 recall events by 72 companies. 5 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Synthes USA HQ, Inc. · 6 events, 8 records
- GE Healthcare, LLC · 4 events, 4 records
- Medtronic Inc. Cardiac Rhythm Disease Management · 3 events, 3 records
- Philips Medical Systems (Cleveland) Inc · 3 events, 3 records
- Siemens Healthcare Diagnostics, Inc. · 3 events, 3 records
- Siemens Medical Solutions USA, Inc · 3 events, 3 records
- Cooper Surgical, Inc. · 2 events, 10 records
- Biomerieux Inc · 2 events, 4 records
FDA root cause
- Software design · 19
- Process control · 15
- Nonconforming Material/Component · 15
- Device Design · 14
- Packaging process control · 10
- Package design/selection · 10
- Employee error · 6
- Mixed-up of materials/components · 5
Class I recalls (5)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1014-2013 · Smiths Medical ASD, Inc. · posted Mar 29, 2013 · Terminated
In rare instances, the BCI Remote Alarm Cables (BCI Cable) are not transferring alarms when used with some remote nurse alarm systems.
- Z-0955-2013 · Vascular Solutions, Inc. · posted Mar 20, 2013 · Terminated
Risk that air may be introduced into the device which may lead to an air embolism.
- Z-0968-2013 · Del Mar Reynolds Medical, Ltd. · posted Mar 20, 2013 · Terminated
A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.
- Z-0940-2013 · Symbios Medical Products, LLC · posted Mar 13, 2013 · Terminated
Received 2 complaints out of 458 distributed pumps in this lot, where flow restrictor bead became displaced, which were reported to permit fast flow of contents.
- Z-0892-2013 · Ad-Tech Medical Instrument Corporation · posted Mar 2, 2013 · Terminated
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall…
Class II recalls (102)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1026-2013 · Advanced Sterilization Products · posted Mar 30, 2013 · Terminated
The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.
- Z-1027-2013 · Advanced Sterilization Products · posted Mar 30, 2013 · Terminated
The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.
- Z-1024-2013 · Zyno Medical LLC · posted Mar 29, 2013 · Terminated
Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump
- Needle Counters ↗Class IIZ-0978-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-0979-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-0980-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Needle Counters ↗Class IIZ-0981-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-0982-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Needle Counters ↗Class IIZ-0983-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Needle Counters ↗Class IIZ-0984-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-0985-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-0986-2013 · Cardinal Health · posted Mar 28, 2013 · Terminated
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
- Z-1015-2013 · CareFusion Corporation · posted Mar 28, 2013 · Terminated
The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
- Z-1016-2013 · Synthes USA HQ, Inc. · posted Mar 28, 2013 · Terminated
The Dens Instrument Set has been designed to facilitate the anterior internal fixation of the Dens (odontoid process). There are currently no screws specifically indicated to treat Dens fractures via anterior screw fixation therefore the affected product…
- Z-1017-2013 · Medtronic Inc. Cardiac Rhythm Disease Management · posted Mar 28, 2013 · Terminated
Medtronic has identified an issue with a specific lot of the Archer 0.035 inch (0.89mm) Super Stiff Guidewire, model ARCSJ200W, where the PTFE (Teflon) coating has the potential to separate from the wire during in-vivo use.
- Z-1018-2013 · Cooper Surgical, Inc. · posted Mar 28, 2013 · Terminated
Incorrect size was printed on the outer product packaging.
- Z-1019-2013 · Medtronic Inc. Cardiac Rhythm Disease Management · posted Mar 28, 2013 · Terminated
Medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, where a change in the manufacturing process has been identified as the cause for…
- Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). ↗Class IIZ-1021-2013 · Microbiologics Inc · posted Mar 28, 2013 · Terminated
Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may…
- Z-1022-2013 · Genetic Testing Institute,inc · posted Mar 28, 2013 · Terminated
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay…
- Z-1023-2013 · Biosense Webster, Inc. · posted Mar 28, 2013 · Terminated
Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.
- Z-1002-2013 · Vital Images, Inc. · posted Mar 27, 2013 · Terminated
Vital Images has found a potential error in measurement generated through the software which affects all versions of Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and…
- Z-1013-2013 · Biomerieux Inc · posted Mar 27, 2013 · Terminated
The product package did not contain the correct version of the package insert listing the limitations of use and performance requirements.
- Z-0995-2013 · Kimberly-Clark Corporation · posted Mar 26, 2013 · Terminated
Some individual container filter units might contain thin areas that may potentially compromise the ability of the filter to maintain a sterile barrier.
- Z-0996-2013 · Smiths Medical ASD, Inc. · posted Mar 26, 2013 · Terminated
An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.
- Z-0997-2013 · Siemens Medical Solutions USA, Inc · posted Mar 26, 2013 · Terminated
The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise…
- Z-0998-2013 · Advanced Sterilization Products · posted Mar 26, 2013 · Terminated
Advanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.
- Z-0987-2013 · Abbott Laboratories, Inc · posted Mar 25, 2013 · Terminated
Abbott has identified that assay files for the ARCHITECT HBsAg Qualitative and HBsAg Qualitative Confirmatory assays (LNs 4P53 and 4P54) on the ARCHITECT i1000SR System were mistakenly made available on abbottdiagnostics.com in October 2012. Additionally,…
- Z-0989-2013 · Biomerieux Inc · posted Mar 25, 2013 · Terminated
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
- Z-0990-2013 · Biomerieux Inc · posted Mar 25, 2013 · Terminated
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
- Z-0991-2013 · Biomerieux Inc · posted Mar 25, 2013 · Terminated
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
- Z-0992-2013 · Siemens Medical Solutions USA, Inc · posted Mar 25, 2013 · Terminated
In the course of the firm's product monitoring activities, a potential risk for patient or operators was determined during the operation of an Artis zeego system with software revision VC14, VC20 and VC21 in conjunction with a specific technical…
- Z-0994-2013 · Toshiba American Medical Systems Inc · posted Mar 25, 2013 · Terminated
Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring…
- Z-0975-2013 · Orthohelix Surgical Designs Inc · posted Mar 22, 2013 · Terminated
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.
- Z-0976-2013 · Becton Dickinson & Co. · posted Mar 22, 2013 · Terminated
Cartridges are not sealing properly during PCR testing and may cause false results, indeterminates and/or invalid runs.
- BS-200 Chemistry Analyzer ↗Class IIZ-0977-2013 · Mindray DS USA, Inc. d.b.a. Mindray North America · posted Mar 22, 2013 · Terminated
An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
- ReClaim Reamer Extension ↗Class IIZ-0972-2013 · DePuy Orthopaedics, Inc. · posted Mar 20, 2013 · Terminated
The tabs on the reamer extension device have broken. DePuy is notifying surgeons of the issue to increase the awareness and to ask surgeons and users to look for tabs before and after surgery to reduce any potential that the tabs are left in the patient.…
- Z-0973-2013 · Keystone Dental Inc · posted Mar 20, 2013 · Terminated
Mislabeled: Cover screw that was in the package was a WD cover screw and not a SD cover screw as labeled
- Z-0974-2013 · Intio Inc · posted Mar 20, 2013 · Terminated
INTIO Inc. distributed a SVM User's Manual and it has an editing mistake that could be interpreted as promoting an unapproved use.
- Z-0970-2013 · AGFA Corp. · posted Mar 19, 2013 · Terminated
The cassettes were shipped with the wrong IP (image plate) size bar code.
- Z-0971-2013 · DePuy Orthopaedics, Inc. · posted Mar 19, 2013 · Terminated
Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to reduce likelihood of potential failure.
- Z-0959-2013 · Allen Medical Systems, Inc. · posted Mar 18, 2013 · Terminated
Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.
- Z-0960-2013 · DePuy Spine, Inc. · posted Mar 18, 2013 · Terminated
Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
- Z-0961-2013 · DePuy Spine, Inc. · posted Mar 18, 2013 · Terminated
Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
- Z-0962-2013 · DePuy Spine, Inc. · posted Mar 18, 2013 · Terminated
Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
- Z-0963-2013 · DePuy Spine, Inc. · posted Mar 18, 2013 · Terminated
Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
- Z-0964-2013 · Stryker Howmedica Osteonics Corp. · posted Mar 18, 2013 · Terminated
The standard Triathlon Femoral Stylus does not fit into the Specialty Sizers and the "R" and "L" markings on the Right Sizer are reversed.
- Z-0965-2013 · Boston Scientific CRM Corp · posted Mar 18, 2013 · Terminated
An internal protective fuse can be unintentionally activated while the device is charging its capacitors for shock delivery or induction. Should this occur, the defibrillator would not be able to deliver therapy or communicate with the Q-TECH Model 2020…
- Z-0967-2013 · Alphatec Spine, Inc. · posted Mar 18, 2013 · Terminated
The rods were incorrectly laser etched as ILLICO Straight CP Tl Rods, 110mm, when they were in fact manufactured to 100mm in length.
- Z-0969-2013 · Molded Products Inc · posted Mar 18, 2013 · Terminated
The recalled product leaks at the connection site.
- Z-0952-2013 · Extremity Medical LLC · posted Mar 15, 2013 · Terminated
The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than…
- Z-0953-2013 · Exactech, Inc. · posted Mar 15, 2013 · Terminated
A dimensional mismatch was identified which can potentially allow for interference between devices.
- Z-0954-2013 · Ebi, Llc · posted Mar 15, 2013 · Terminated
Increased risk of failure at the driver tip during screw insertion or removal.
- Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431.Class IIZ-0956-2013 · Synthes USA HQ, Inc. · posted Mar 15, 2013 · Terminated
It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
- Z-0957-2013 · Siemens Healthcare Diagnostics, Inc. · posted Mar 15, 2013 · Terminated
Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS…
- Z-0958-2013 · Beckman Coulter Inc. · posted Mar 15, 2013 · Terminated
Beckman Coulter is including additional information and instructions in the Creatinine (OSR6x78) IFU and Setting Sheet, relating to calibration stability and quality control frequency to assure product performance.
- Z-0948-2013 · Siemens Medical Solutions USA, Inc · posted Mar 14, 2013 · Terminated
Sporadic hang-up malfunctions concerning 3D reconstructions during activated "Preview Image" functionality.
- TotalCare SpO2RT2 bed. ↗Class IIZ-0949-2013 · Hill-Rom, Inc. · posted Mar 14, 2013 · Terminated
In certain situations a software problem with the TotalCare SpO2RT 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the likelihood of skin irritation for patients…
- Z-0950-2013 · Philips Medical Systems (Cleveland) Inc · posted Mar 14, 2013 · Terminated
This field change order is being released to update software and customer release notes to the affected installed base.
- Vial Label: ↗Class IIZ-0951-2013 · Diagnostica Stago, Inc. · posted Mar 14, 2013 · Terminated
Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.
- Z-0943-2013 · Smith & Nephew Inc · posted Mar 13, 2013 · Terminated
75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.
- Z-0944-2013 · Smith & Nephew Inc · posted Mar 13, 2013 · Terminated
75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.
- Z-0945-2013 · Philips Medical Systems (Cleveland) Inc · posted Mar 13, 2013 · Terminated
Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
- Z-0946-2013 · Natus Neurology Incorporated · posted Mar 13, 2013 · Terminated
Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the on/off button. The overheating may result in discomfort ranging from too hot up to the potential for a…
- Z-0947-2013 · GE Healthcare, LLC · posted Mar 13, 2013 · Terminated
A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.
- Z-0942-2013 · Synthes USA HQ, Inc. · posted Mar 12, 2013 · Terminated
There is the potential for the anterior footplates to not fit onto the Titanium Midface Distractor Assembly as the tab of some distractors may be oversized.
- GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging ↗Class IIZ-0934-2013 · GE Healthcare, LLC · posted Mar 11, 2013 · Terminated
Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
- Z-0935-2013 · GE Healthcare, LLC · posted Mar 11, 2013 · Terminated
An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports.
- Z-0936-2013 · Bacterin International, Inc. · posted Mar 11, 2013 · Terminated
A portion of a silicone wound drain was found to be cracked.
- Z-0937-2013 · Vygon Corporation · posted Mar 11, 2013 · Terminated
Product was packaged with the incorrect introducer needle. The kit label indicates a "safety" needle is included; the kits were packaged with a non-safety needle.
- Z-0938-2013 · Synthes USA HQ, Inc. · posted Mar 11, 2013 · Terminated
Lot number 7542279, of the 45 mm Click 'X Pedicle Screw (Part number 487.072), was incorrectly etched as "Click 'X Pedicle Screw 05.2 L40 TAN violet" although the actual length of the screw is 45 mm.
- Z-0939-2013 · Draeger Medical, Inc. · posted Mar 11, 2013 · Terminated
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a…
- Z-0941-2013 · Synthes USA HQ, Inc. · posted Mar 11, 2013 · Terminated
The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.
- Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1Class IIZ-0931-2013 · Elekta, Inc. · posted Mar 10, 2013 · Terminated
The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
- Z-0932-2013 · Spine Smith Partners LP · posted Mar 10, 2013 · Terminated
Implants were incorrectly laser marked as 23mm instead of 28mm.
- Z-0929-2013 · Abbott Medical Optics Inc (AMO) · posted Mar 9, 2013 · Terminated
Two production orders of these IOLs that contain mislabeled IOLs.
- Z-0930-2013 · Siemens Medical Solutions USA, Inc. · posted Mar 9, 2013 · Terminated
On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.
- Command Instrument Broach Handle. Used in Total Hip Arthroplasty. Catalog Number: 6266-0-100 ↗Class IIZ-0919-2013 · Stryker Howmedica Osteonics Corp. · posted Mar 8, 2013 · Terminated
Reports of handle fracture and subsequent disassociation of cam assembly subcomponents during use.
- Z-0920-2013 · Integra LifeSciences Corp. · posted Mar 8, 2013 · Terminated
There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue.…
- Z-0921-2013 · Genesys Orthopedic Systems, LLC · posted Mar 8, 2013 · Terminated
Manufacturing residue may be present on the cervical screws. Use of this product may result in patient infection, delayed fusion or non-fusion (pseudarthrosis), inflammatory response, or revision surgery.
- Z-0922-2013 · Instratek, Incorporated · posted Mar 8, 2013 · Terminated
Instratek, Inc.has received multiple complaints of the 1052 Trigger Finger Release Blades which could not be advanced into the cannula through which it is used during trigger finger release procedures.
- Z-0923-2013 · Keystone Dental Inc · posted Mar 8, 2013 · Terminated
Mislabeled: Genesis Straight Esthetic 5.0mm Contour ,3mm cuff labeled as Genesis Straight Esthetic 4.0 mm, 1 mm cuff abutment,
- Z-0924-2013 · Keystone Dental Inc · posted Mar 8, 2013 · Terminated
Mislabeled: Genesis Straight Esthetic 4.0mm Contour ,1mm cuff labeled as Genesis Straight Esthetic 5.0 mm, 3mm cuff abutment
- Z-0925-2013 · Biomet, Inc. · posted Mar 8, 2013 · Terminated
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not…
- Medtronic O-arm Imaging System ↗Class IIZ-0926-2013 · Medtronic Navigation, Inc. · posted Mar 8, 2013 · Terminated
Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
- Z-0927-2013 · Innominata Dba Genbio · posted Mar 8, 2013 · Terminated
Multiple incidents of low-reacting assay plates.
- Z-0928-2013 · Sorin Group USA, Inc. · posted Mar 8, 2013 · Terminated
Isoline defibrillation leads distributed may have internal insulation breach.
- Z-0918-2013 · Ethicon Endo-Surgery Inc · posted Mar 7, 2013 · Terminated
Damage to the packaging may have compromised the sterility of the device.
- Z-0909-2013 · Bard Access Systems · posted Mar 6, 2013 · Terminated
Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong size introducer being included with…
- Z-0910-2013 · Bard Access Systems · posted Mar 6, 2013 · Terminated
Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong size introducer being included with…
- Z-0912-2013 · Atricure Inc · posted Mar 6, 2013 · Terminated
Products from lot no. 41865 may have been sent out with Incorrect Instructions for Use (IFU) labeling.
- Z-0913-2013 · Beckman Coulter Inc. · posted Mar 6, 2013 · Terminated
Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.
- Z-0914-2013 · Churchill Medical Systems, Inc. · posted Mar 6, 2013 · Terminated
The label on the device states that the item number is CLGY-2210 when the item number on the device label should read CLGY-2010. This is a 20Ga Huber needle set; however, the product label states the pouch contains a 22Ga set.
- Z-0915-2013 · Steris Corporation · posted Mar 6, 2013 · Terminated
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced,…
- Philips Medical System Extended Brilliance Workspace (EBW), Computed Tomography X-Ray System. ↗Class IIZ-0906-2013 · Philips Medical Systems (Cleveland) Inc · posted Mar 5, 2013 · Terminated
Incorrect values may be displayed. Philips is issuing a software update (version 4.5.5) to the Extended Brilliance Workspace (EBW) systems currently running software version(s) 4.0.2, V4.5.2 or V4.5.3, addressing a number of non-conformances.
- Z-0907-2013 · Siemens Medical Solutions USA, Inc · posted Mar 5, 2013 · Terminated
Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus…
- Medtronic DPL Single Stage Venous Cannula with Metal Tip, used in cardiopulmonary bypass procedures.Class IIZ-0908-2013 · Medtronic Inc. Cardiac Rhythm Disease Management · posted Mar 5, 2013 · Terminated
Potential for uncharacteristically rough metal edges to occur along the inner diameter of the metal tip side ports of the cannula.
- Philips Digital Diagnost ↗Class IIZ-0900-2013 · Philips Healthcare Inc. · posted Mar 1, 2013 · Terminated
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates…
- Z-0901-2013 · GE Healthcare, LLC · posted Mar 1, 2013 · Terminated
Incorrect views and measurements possible. GE Healthcare became aware of multiple potential safety issues involving the Advantage workstation, Volume Viewer Plus and Volume Analysis.
- Z-0902-2013 · Salyer Prn Biomedical · posted Mar 1, 2013 · Terminated
Concerns that the addition of the UPS Battery Backup to the Century Console Base Shelf did not comply with the FDA premarket notification regulations.
- Z-0903-2013 · Synthes USA HQ, Inc. · posted Mar 1, 2013 · Terminated
It was discovered that the sheer pin on the Multi-Vector Distractor may not be made according to product specifications.
- Synthes Multi-Vector Distractor, Multi-axial Mandible Distractor (Right), Part Number 487.966. ↗Class IIZ-0904-2013 · Synthes USA HQ, Inc. · posted Mar 1, 2013 · Terminated
It was discovered that the sheer pin on the Multi-Vector Distractor may not be made according to product specifications.
- Synthes Multi-Vector Distractor - Multi-axial Mandible Distractor (Left), Part Number 487.967. ↗Class IIZ-0905-2013 · Synthes USA HQ, Inc. · posted Mar 1, 2013 · Terminated
It was discovered that the sheer pin on the Multi-Vector Distractor may not be made according to product specifications.
Class III recalls (17)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Dual Tip Marker, Sterile ↗Class IIIZ-1025-2013 · Viscot Medical LLC · posted Mar 29, 2013 · Terminated
Dual Tip Marker has inconsistent ink flow on the black permanent ink side.
- Z-1004-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1005-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1006-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1007-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1008-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1009-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1010-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1011-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1012-2013 · Cooper Surgical, Inc. · posted Mar 27, 2013 · Terminated
Pessary mislabeled outer package may not reflect the size of the product contained in the package
- Z-1001-2013 · CareFusion 213, LLC · posted Mar 26, 2013 · Terminated
An internal review of CareFusion's labeling for preoperative skin prep product codes was recently conducted. The labeling of the product codes identified above state that the contents of the package are sterile and does not clarify that this sterility…
- Z-0988-2013 · Siemens Healthcare Diagnostics, Inc. · posted Mar 25, 2013 · Terminated
Correlation slope bias and accuracy shift
- Z-0966-2013 · Volk Optical Inc · posted Mar 18, 2013 · Terminated
The defect was with mis-engraving of the lot # directly on the product and its packaging. Mixed up device engraving of V20LC (lot # BB09213) with that from V60C (lot # BB09205).
- Z-0933-2013 · Siemens Healthcare Diagnostics, Inc. · posted Mar 11, 2013 · Terminated
The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim…
- Z-0911-2013 · Becton Dickinson & Company · posted Mar 6, 2013 · Terminated
BD is conducting a recall of the BD Vacutainer Plus Plastic Plasma Separator Tube (PST) with Light Green/BD Homogard Closure due to a deformed top of the tube reservoir reported in two customer complaints.
- D3000 Advantage Drive System, 115v ↗Class IIIZ-0916-2013 · Linvatec Corp. dba ConMed Linvatec · posted Mar 6, 2013 · Terminated
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment:…
- Z-0917-2013 · Linvatec Corp. dba ConMed Linvatec · posted Mar 6, 2013 · Terminated
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment:…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.