Trilogy F/M Acetabular Shell with Uni-Hole recallZ-0771-2013 · Zimmer, Inc. · initiated Sep 14, 2012
Class II Terminated All recalls posted in February 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
FDA root cause category: Process control
What the company is doing
Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.
Recall details
- Recall number
- Z-0771-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 14, 2012
- Posted by the FDA
- Feb 4, 2013
- Recalling firm
- Zimmer, Inc.
- Quantity
- 275,952 units
- Product code
- LPH
- FDA submissions
- K934765
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
Product description
Trilogy F/M Acetabular Shell with Uni-Hole. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
Code information (models, lots, serial numbers)
61282267 61633887 61260645 61237937 61500763 61397941 61584769 61199849 61429445 61412089 00369745
Other devices under the same FDA submission
8 devices cleared under K934765 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy® 6200-56-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Trilogy® 6200-50-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Trilogy® 6200-52-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-60-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-54-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-62-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-66-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6105-50-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 63531).
- Z-0769-2013 · TerminatedTrilogy F/M Acetabular Shell without holes. Several ODs: 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM. The Trilogy Acetabular Sy…
- Z-0770-2013 · TerminatedTrilogy F/M Acetabular Shell without holes SOLID HA. Several ODs: 38MM, 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM. The Trilogy…
- Z-0772-2013 · TerminatedTrilogy Acetabular Shell Spiked. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
- Z-0773-2013 · TerminatedTrilogy Acetabular Shell with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating …
- Z-0775-2013 · TerminatedTrilogy Acetabular Shells with Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip dam…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page