Medical device recalls: February 2013
The FDA posted 139 medical device recall records in February 2013, from 83 recall events by 63 companies. 8 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 6 events, 8 records, 2 Class I
- Philips Medical Systems (Cleveland) Inc · 4 events, 7 records
- Remel Inc · 4 events, 5 records
- Philips Healthcare Inc. · 4 events, 4 records
- Ultradent Products, Inc. · 2 events, 3 records
- Aesculap, Inc. · 2 events, 2 records
- Elekta, Inc. · 2 events, 2 records
- Integra LifeSciences Corp. · 2 events, 2 records
FDA root cause
- Component design/selection · 31
- Software design · 18
- Process control · 17
- Nonconforming Material/Component · 15
- Device Design · 13
- Packaging · 6
- Material/Component Contamination · 6
- Unknown/Undetermined by firm · 4
Class I recalls (8)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- LPS Diaphyseal SleevesClass IZ-0828-2013 · DePuy Orthopaedics, Inc. · posted Feb 15, 2013 · Terminated
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to…
- LPS Diaphyseal SleevesClass IZ-0829-2013 · DePuy Orthopaedics, Inc. · posted Feb 15, 2013 · Terminated
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to…
- LPS Diaphyseal SleevesClass IZ-0830-2013 · DePuy Orthopaedics, Inc. · posted Feb 15, 2013 · Terminated
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to…
- LPS Diaphyseal SleevesClass IZ-0831-2013 · DePuy Orthopaedics, Inc. · posted Feb 15, 2013 · Terminated
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to…
- Z-0795-2013 · Lumenis Limited · posted Feb 14, 2013 · Terminated
Lumenis Ltd is recalling certain models of VersaCut Morcellator devices to perform corrective labeling as a result of three reported events of air embolism secondary to incorrect reversal tubing system hook-up.
- Z-0763-2013 · St Jude Medical Inc · posted Feb 11, 2013 · Terminated
The distal end of the core wire of the TorqVue FX Delivery System could potentially fracture when exposed to a combination of certain cardiac anatomies and usage conditions.
- Z-0748-2013 · GE Healthcare, LLC · posted Feb 8, 2013 · Terminated
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
- Z-0749-2013 · GE Healthcare, LLC · posted Feb 8, 2013 · Terminated
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Class II recalls (125)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0893-2013 · Stryker Spine · posted Feb 28, 2013 · Terminated
Surgeons have experienced unthreading of the screwdrivers outer shafts during use.
- Z-0894-2013 · Advanced Neuromodulation Systems Inc. · posted Feb 28, 2013 · Terminated
Unintended Fluid ingress into the device header has been shown.
- Z-0895-2013 · Orthofix, Inc · posted Feb 28, 2013 · Terminated
There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the…
- Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700).Class IIZ-0896-2013 · Orthofix, Inc · posted Feb 28, 2013 · Terminated
There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to…
- Z-0897-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 28, 2013 · Terminated
The system may not default to the same gating trigger with which the data were acquired. This could cause a risk for patients.
- Z-0898-2013 · Medical Components, Inc dba MedComp · posted Feb 28, 2013 · Terminated
The affected product was packaged with the incorrect introducer needle. The kit label indicates a "safety" needle is included; the kits were packaged with a non-safety needle.
- Z-0899-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 28, 2013 · Terminated
The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.
- Z-0703-2013 · GE Healthcare, LLC · posted Feb 27, 2013 · Terminated
When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The…
- Z-0704-2013 · GE Healthcare, LLC · posted Feb 27, 2013 · Terminated
When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The…
- Z-0891-2013 · Myco Medical Supplies Inc · posted Feb 27, 2013 · Terminated
Pouch labeling/printing errors and defective packaging resulting in possible compromised sterility.
- Z-0761-2013 · GE Healthcare, LLC · posted Feb 26, 2013 · Terminated
GE Healthcare has become aware of a software issue on the interface of the Brivo XR385 DR-F/ Model 5271997 radiographic product related to Visual indication as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-1-3.
- Z-0786-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 26, 2013 · Terminated
The low contrast pins in the Image Quality (IQ) phantom deteriorates over time. This prevents the customers from doing the Daily Image Quality Checks as defined in the current Instructions for Use (IfU).
- Z-0879-2013 · Baylis Medical Company Inc. · posted Feb 26, 2013 · Terminated
A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse, the pouch package of 5 of the 10 individually packaged cannula had not been sealed along one edge. Thus the pouch was open and the device could…
- Z-0880-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 26, 2013 · Terminated
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm…
- Z-0881-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 26, 2013 · Terminated
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm…
- Z-0882-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 26, 2013 · Terminated
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm…
- Z-0883-2013 · Philips Medical Systems (Cleveland) Inc · posted Feb 26, 2013 · Terminated
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm…
- Z-0884-2013 · Philips Healthcare Inc. · posted Feb 26, 2013 · Terminated
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates…
- Z-0885-2013 · Elekta, Inc. · posted Feb 26, 2013 · Terminated
If the Apex collimator is incorrectly installed on the radiation head, a touchguard inhibit activates the view restrictions dialog box only shows the top 12 restrictions.
- Z-0886-2013 · Integra LifeSciences Corp. · posted Feb 26, 2013 · Terminated
Addendum to the Instructions for Use. The addendum emphasizes the warnings and precautions already contained in the device's IFU and also provides additional warnings and precautions that are applicable to the use of electrosurgical equipment in health…
- Z-0887-2013 · Aesculap, Inc. · posted Feb 26, 2013 · Terminated
The bipolar energy did not stop after release of the foot pedal.
- Z-0888-2013 · Cellestis Inc · posted Feb 26, 2013 · Terminated
Cellestis is recalling QuantiFERON - TB Gold TB Blood Collection Tubes and HA TB Blood Collection Tubes due to potential contamination.
- Z-0889-2013 · Cellestis Inc · posted Feb 26, 2013 · Terminated
Cellestis is recalling QuantiFERON - TB Gold TB Blood Collection Tubes and HA TB Blood Collection Tubes due to potential contamination.
- Z-0890-2013 · Oculus Innovative Sciences Inc · posted Feb 26, 2013 · Terminated
Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.
- Z-0877-2013 · Intel-GE Care Innovations LLC · posted Feb 23, 2013 · Terminated
Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.
- Versafitcup Impacting RingClass IIZ-0878-2013 · Medacta Usa Inc · posted Feb 23, 2013 · Terminated
Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to…
- Z-0873-2013 · Mc-NEIL-PPC, Inc. · posted Feb 22, 2013 · Terminated
This product is not cleared for marketing in the US.
- Z-0874-2013 · Mc-NEIL-PPC, Inc. · posted Feb 22, 2013 · Terminated
This product is not cleared for marketing in the US.
- Z-0875-2013 · Mc-NEIL-PPC, Inc. · posted Feb 22, 2013 · Terminated
This product is not cleared for marketing in the US.
- Z-0869-2013 · Integra LifeSciences Corp. · posted Feb 21, 2013 · Terminated
Integra determined there is a possibility that the Base Units may break during use.
- Z-0870-2013 · Medrad Inc · posted Feb 21, 2013 · Terminated
The seal between the white polyethylene and the clear plastic may be missing on the Hand Controller Sheath package, resulting in a potential breach of the sterility of this accessory.
- Z-0871-2013 · Philips Healthcare Inc. · posted Feb 21, 2013 · Terminated
Philips HeartStart XL+ Defibrillator/Monitor may become locked out of clinical use
- Z-0872-2013 · Hill-Rom, Inc. · posted Feb 21, 2013 · Terminated
Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest…
- Z-0838-2013 · Ansell Healthcare Products LLC · posted Feb 20, 2013 · Terminated
Out of an abundance of caution, Ansell is implementing a voluntary recall of the Zero Latex Condom (regular) product. The condom has met or exceeded test and quality standards required by the Food and Drug Administration. However, we are experiencing a…
- Z-0839-2013 · Maquet Cardiovascular Us Sales, Llc · posted Feb 20, 2013 · Terminated
ELS 15 Fr Cannula could potentially contain an occluded arterial lumen.
- Z-0840-2013 · Stryker Sustainability Solutions · posted Feb 20, 2013 · Terminated
Reports that the AC Power Adaptor may not operate due to wires coming loose in the unit.
- Monaco RTP System. ↗Class IIZ-0841-2013 · Elekta, Inc. · posted Feb 20, 2013 · Terminated
The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
- Z-0842-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of packaging integrity
- Z-0843-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0844-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0845-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0846-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0847-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0848-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0849-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0850-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0851-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0852-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0853-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0854-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0855-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0856-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0857-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0858-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0859-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0860-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0861-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0862-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0863-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0864-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0865-2013 · Codman & Shurtleff, Inc. · posted Feb 20, 2013 · Terminated
Sterility of device may be compromised due to lack of package integrity
- Z-0866-2013 · Merit Medical Systems, Inc. · posted Feb 20, 2013 · Terminated
A consumer complaint that a catheter was observed to be in several pieces under patient fluoroscopic visualization.
- Z-0867-2013 · Varian Medical Systems, Inc. Oncology Systems · posted Feb 20, 2013 · Terminated
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
- Z-0868-2013 · Vascular Solutions, Inc. · posted Feb 20, 2013 · Terminated
Vascular Solutions, Inc has discovered a potential problem with a number of lots of Supercross microcatheters. A potential for surface contamination on aged catheters was observed during in-house testing.
- Z-0837-2013 · International Technidyne Corp. · posted Feb 19, 2013 · Terminated
Some Pro Time3 Test Cuvettes within a specified lot range may recover lower than expected Prothrombin Time/International Normalized Ratio (PT/INR) results.
- Z-0825-2013 · Healthtronics, Inc. · posted Feb 15, 2013 · Terminated
The Endocare 3.8mm RenalCryo Cryoprobe is being recalled because the label on the outer box has the incorrect expiration date.
- Z-0826-2013 · ConMed Corporation · posted Feb 15, 2013 · Terminated
Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.
- Z-0827-2013 · ConMed Corporation · posted Feb 15, 2013 · Terminated
Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.
- Z-0832-2013 · Carefusion 303 Inc · posted Feb 15, 2013 · Terminated
CareFusion is recalling the SmartSite Gravity Blood Set as a result of a potential for separation at the connection of the drip chamber and tubing . The separation may be observed at priming or observed during use. If a separation between the drip chamber…
- Z-0835-2013 · Greiner Bio-One North America, Inc. · posted Feb 15, 2013 · Terminated
The VACUETTE Safety Blood Collection Set + Luer Adapter may display leaking (small droplets) during use due to a crack on the luer connector piece.
- Z-0654-2013 · Lockheed Martin Gyrocam Systems, Inc. · posted Feb 14, 2013 · Terminated
These camera systems were manufactured by LMGS and found to fail to comply with the FDA laser performance standard during an FDA inspection. These camera systems incorporate Class IIIb surveying, leveling, and alignment lasers and were entered into US…
- Z-0814-2013 · GE Healthcare, LLC · posted Feb 14, 2013 · Terminated
GE has become aware of multiple issues affecting the Dash 3000/4000/5000 and DashPort 2 monitor. This report of correction encompasses the DashPort 2 device, in which the DashPort 2 stops communicating with the Dash, resulting in Loss of Monitoring at the…
- Z-0815-2013 · NAImco Inc dba Richmar Inc · posted Feb 14, 2013 · Terminated
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
- Z-0816-2013 · NAImco Inc dba Richmar Inc · posted Feb 14, 2013 · Terminated
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
- Z-0817-2013 · NAImco Inc dba Richmar Inc · posted Feb 14, 2013 · Terminated
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
- Z-0818-2013 · NAImco Inc dba Richmar Inc · posted Feb 14, 2013 · Terminated
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
- Z-0819-2013 · Quidel Corporation · posted Feb 14, 2013 · Terminated
Quidel Corporation has initiated a voluntary recall of specific lots of the Sofia Influenza A+B FIA kits because of false positive results which may lead to misdiagnosis.
- Z-0820-2013 · Viterion TeleHealthcare Llc · posted Feb 14, 2013 · Terminated
Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04…
- Z-0821-2013 · Viterion TeleHealthcare Llc · posted Feb 14, 2013 · Terminated
Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04…
- Z-0822-2013 · Viterion TeleHealthcare Llc · posted Feb 14, 2013 · Terminated
Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04…
- Z-0823-2013 · GE Healthcare, LLC · posted Feb 14, 2013 · Terminated
GE Healthcare has become aware of potential issues associated with the CARESCAPE Monitor B650. There have been no reported illnesses or injuries that have occurred with the use of the device.
- Z-0824-2013 · Philips Healthcare Inc. · posted Feb 14, 2013 · Terminated
VS2+ screen may lock up and stop monitoring during patient use
- Z-0809-2013 · Chenica Inc. · posted Feb 13, 2013 · Terminated
The wheelchairs may not meet the specifications of WC-19 (crash testing), therefore the WC-19 verbiage and corresponding logo are removed from the flyers on the Gunnell Rehab Series of Wheelchairs (GRSW).
- Z-0810-2013 · Mizuho OSI · posted Feb 13, 2013 · Terminated
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
- GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Physiological Patient Monitor and accessories. ↗Class IIZ-0813-2013 · GE Healthcare, LLC · posted Feb 13, 2013 · Terminated
GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor. These issues may occur when the Ethernet controller in the XPC823 CPU on the Dash and DashPort 2 motherboard functions intermittently. GE Healthcare has identified…
- Z-0802-2013 · Terumo Cardiovascular Systems Corporation · posted Feb 12, 2013 · Terminated
Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
- Z-0803-2013 · Breg Inc · posted Feb 12, 2013 · Terminated
The recall was initiated because Breg has determined that some Sterile Polar Pads products manufactured from January 2010 to October 2012 may have sustained damage to the product packaging which may compromise product sterility assurance.
- BREG Sterile Polar Pads ↗Class IIZ-0804-2013 · Breg Inc · posted Feb 12, 2013 · Terminated
The recall was initiated because Breg has determined that some Sterile Polar Pads products manufactured from January 2010 to October 2012 may have sustained damage to the product packaging which may compromise product sterility assurance.
- Z-0805-2013 · Aesculap, Inc. · posted Feb 12, 2013 · Terminated
The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in…
- Z-0806-2013 · Remel Inc · posted Feb 12, 2013 · Terminated
The firm is recalling the products due to a potential for false negative test results.
- Z-0807-2013 · Remel Inc · posted Feb 12, 2013 · Terminated
The firm is recalling the products due to a potential for false negative test results.
- Z-0808-2013 · Remel Inc · posted Feb 12, 2013 · Terminated
The firm is recalling the product because the cartridges may contain discs that have an insufficient dose of antibiotic.
- Z-0798-2013 · Mindray DS USA, Inc. d.b.a. Mindray North America · posted Feb 11, 2013 · Terminated
Mindray has identified a software anomaly that may where, in the presence of an asystole, the Telepack ECG waveform may produce a unique step pattern (instead of a flat line) if Lead 1 or V is being utilized for detecting cardiac arrhythmia.
- Z-0799-2013 · Fresenius Medical Care Holdings, Inc. · posted Feb 11, 2013 · Terminated
There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.
- Z-0800-2013 · Bacterin International, Inc. · posted Feb 11, 2013 · Terminated
Expired product was shipped to two customers.
- Z-0801-2013 · Remel Inc · posted Feb 11, 2013 · Terminated
The product may contain low level microbial contamination which could result in incorrect results.
- Z-0791-2013 · Medacta Usa Inc · posted Feb 7, 2013 · Terminated
The tip of the Quadra Trial Broach size 0, broke during a surgery.
- SeriScaffold Surgical Scaffold ↗Class IIZ-0794-2013 · Allergen Medical · posted Feb 7, 2013 · Terminated
Product is dual packaged in an inner and outer pouch, the outer pouch seal may be compromised and sterility may not be assured
- Custom Breast Biopsy Tray, Catalog # B 9-16261, packaged in bags, 1 tray/bag, 20 bags/case. ↗Class IIZ-0796-2013 · Custom Healthcare Systems, Inc. · posted Feb 7, 2013 · Terminated
Surgical procedure trays are packed with pharmaceuticals for injection that may contain particulate matter.
- Z-0797-2013 · Custom Healthcare Systems, Inc. · posted Feb 7, 2013 · Terminated
Surgical procedure trays are packed with pharmaceuticals for injection that may contain particulate matter.
- Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK,… ↗Class IIZ-0790-2013 · Remel Inc · posted Feb 6, 2013 · Terminated
The firm is recalling the product due to some cartridges may contain individual discs that are not impregnated with antibiotic. The firm notified their customers of the recall with a notification letter and requested a confirmation fax back of receipt.
- Z-0787-2013 · Exactech, Inc. · posted Feb 5, 2013 · Terminated
Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was inadvertently transposed.
- Z-0788-2013 · Exactech, Inc. · posted Feb 5, 2013 · Terminated
Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was inadvertently transposed.
- Z-0789-2013 · Linvatec Corp. dba ConMed Linvatec · posted Feb 5, 2013 · Terminated
ConMed Linvatech recalled one lot of their PopLok bone Punch 3.5mm because the metal shaft diameter is 4.5mm when it should be 3.5mm.
- Z-0769-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0770-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0771-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Trilogy Acetabular Shell Spiked.Class IIZ-0772-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0773-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0774-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0775-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- Z-0776-2013 · Zimmer, Inc. · posted Feb 4, 2013 · Terminated
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy…
- PREMIER LIFECARE RECLINER 5400 ↗Class IIZ-0777-2013 · Winco Mfg., LLC · posted Feb 4, 2013 · Terminated
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
- Z-0778-2013 · Winco Mfg., LLC · posted Feb 4, 2013 · Terminated
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
- Z-0779-2013 · Winco Mfg., LLC · posted Feb 4, 2013 · Terminated
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
- Z-0780-2013 · Winco Mfg., LLC · posted Feb 4, 2013 · Terminated
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
- Z-0781-2013 · Synthes USA HQ, Inc. · posted Feb 4, 2013 · Terminated
Although the reported incidence is low, there is the potential for the drill stop to slip on the drill bit resulting in an adverse event.
- Z-0782-2013 · Synthes USA HQ, Inc. · posted Feb 4, 2013 · Terminated
Although the reported incidence is low, there is the potential for the drill stop to slip on the drill bit resulting in an adverse event.
- Z-0783-2013 · The Anspach Effort, Inc. · posted Feb 4, 2013 · Terminated
Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did…
- Z-0784-2013 · Lansinoh Laboratories Inc · posted Feb 4, 2013 · Terminated
The Spanish translation on the Instructions for Use (IFU) insert and product package for hot and cold packs has incorrect heating instructions which would cause product to over heat.
- Z-0785-2013 · Invacare Corporation · posted Feb 4, 2013 · Terminated
The potential exists for the wheel to rotate freely despite engagement of hub brake.
- Z-0768-2013 · Oxus America, Inc. · posted Feb 2, 2013 · Terminated
Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.
- Z-0765-2013 · Ormco/Sybronendo · posted Feb 1, 2013 · Terminated
The length markings on the Lightspeed LSX files were wrong.
- Z-0766-2013 · Tobii Assistive Technology, Inc. · posted Feb 1, 2013 · Terminated
Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.
- Z-0767-2013 · Philips Healthcare Inc. · posted Feb 1, 2013 · Terminated
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Class III recalls (6)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0876-2013 · Ultradent Products, Inc. · posted Feb 22, 2013 · Terminated
Orthodontic brackets distributed was incorrectly labeled.
- Z-0836-2013 · Animas Corporation · posted Feb 17, 2013 · Terminated
Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display.
- Z-0811-2013 · Ultradent Products, Inc. · posted Feb 13, 2013 · Terminated
Incorrect packaging of regular set and fast set cartridges.
- Z-0812-2013 · Ultradent Products, Inc. · posted Feb 13, 2013 · Terminated
Incorrect packaging of regular set and fast set cartridges.
- Z-0792-2013 · Diagnostic Hybrids Inc · posted Feb 7, 2013 · Terminated
RhMK product fungal contamination.
- OPTIGEN Food 30 ↗Class IIIZ-0793-2013 · Hitachi Chemical Diagnostics Inc · posted Feb 7, 2013 · Terminated
There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.