Sol Sys 8in Calc 1.5/15.0 recallZ-0575-2013 · DePuy Orthopaedics, Inc. · initiated Dec 3, 2012
Class II Terminated All recalls posted in December 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
FDA root cause category: Device Design
What the company is doing
Deputh Orthopaedics sent an Urgent Information - Medical Device Recall Notice dated December , 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The notification letter instructed consignees to please cease further distribution or use of products immediately. The purpose of the communication was to inform them of this recall and request acknowledgement of receipt of this letter by their facility. Reconciliation forms should be returned to your DePuy Sales Representative or faxed to the following Fax Number: 574-371-4939. For questions regarding this recall call 574-267-8143.
Recall details
- Recall number
- Z-0575-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 3, 2012
- Posted by the FDA
- Dec 20, 2012
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 1732 total all product codes
- Product code
- LPH
- FDA submissions
- K060581
- Distribution
- Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
Product description
SOL SYS 8IN CALC 1.5/15.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Product Code: 157215150 and Lot codes: 111620, DD3BR1000, 126364, DD3BRA000, 134668A, DG1N91000, A37A71000, DN5H11000, A37A7A000, DN5H1B000, A37A7C, DT6ED1000, A3RBC1000, E1RHJ1000, A3RBCA000, E4HSE1000, A3YAV1000, E4LLG1000, A4VAK1000, E5EB41000, A6ACH1000, E5EB4A000, A88AA1000, E5WJ81000, AP2BB1000, E73B51, AV3DP1000, EC3DN1000, B2LAB1000, ED5KV1000, B3FAE1000, EJ4EG1000, B46AH1000, ER8EY1000, B54FC1000, EW6G31000, BS2GG1000, EY2ET1000, BY3BD1000, FB3E21, C1PC91000, FB8P81, C34B71000, FD4SX1, C3KA31000, FF8DT1, C4JCP1000, FG7DP1, C4VCP1000, FJ9L91, C5FH71000, X32EY1000, C5FH7A, X32EYA000, CC5JP1000, X32EYB000, CE6FL1000, X49FD1000, CH7EF1000, X49FDA000, CL5PE1000, X4ABK1000, CL5PE1A, X4ABKA000, CV5BS1000, X4ABKB000, CV5BSA000, X4GD91000, CX8CV1000, YA6GR1000, CX8CVA000, YF5DJ1000, CY7B11000, YH2C21000, D1TLF1000, YH2C2A000, D32J51000, Z41AS1000, D32J5A000, Z41ASA000, D37GC1000, Z5TH71000, D46LF1000, Z6BB81000, D52LX1000, Z6BB8A000, D81NN1000, Z83A11000, DA2EM1000, ZN7HS1000, DA2EMA000, ZN7HSA000, DA2EMB000, ZN7HSB000, and DA8MD1000.
Other devices under the same FDA submission
10 devices cleared under K060581 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SOLUTION SYSTEM 1572-01-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- SOLUTION SYSTEM 1572-02-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- SOLUTION SYSTEM 1572-11-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- SOLUTION SYSTEM 1572-01-106Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- SOLUTION SYSTEM 1572-02-121Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- SOLUTION SYSTEM 1572-01-121Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- SOLUTION SYSTEM 1572-02-106Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- SOLUTION SYSTEM 1572-08-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- SOLUTION SYSTEM 1572-10-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- SOLUTION SYSTEM 1572-07-136Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 63545).
- Z-0573-2013 · TerminatedSOL SYS 8IN CALCAR SZ12 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sle…
- Z-0574-2013 · TerminatedSOL SYS F 8 CALC 13.5MM LG Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam …
- Z-0576-2013 · TerminatedSOL SYS 8IN CALC 1.5/16.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam s…
- Z-0577-2013 · TerminatedSOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam s…
- Z-0578-2013 · TerminatedSOL SYS 8IN CALC 1.5/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam s…
- Z-0579-2013 · TerminatedSOL SYS F 9 CALC L 13.5MM LG Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0580-2013 · TerminatedSOL SYS L 9IN CALC 2.25/15.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0581-2013 · TerminatedSOL SYS L 9IN CALC 2.25/16.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0582-2013 · TerminatedSOL SYS L 9IN CALC 2.25/18 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam …
- Z-0583-2013 · TerminatedSOL SYS L 9IN CALC 2.25/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0584-2013 · TerminatedSOL SYS L 9IN CALC 2.25/21 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam …
- Z-0585-2013 · TerminatedSOL SYS L 9IN CALC 2.25/22.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0586-2013 · TerminatedSOL SYS F 9 CALC L 13.5MM LG Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0587-2013 · TerminatedSOL SYS R 9IN CALC 2.25/15.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0588-2013 · TerminatedSOL SYS R 9IN CALC 2.25/16.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0589-2013 · TerminatedSOL SYS R 9IN CALC 2.25/18 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam …
- Z-0590-2013 · TerminatedSOL SYS R 9IN CALC 2.25/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
- Z-0591-2013 · TerminatedSOL SYS R 9IN CALC 2.25/21 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam …
- Z-0592-2013 · TerminatedSOL SYS R 9IN CALC 2.25/22.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foa…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page