Skip to content

Sol Sys R 9in Calc 2.25/19.5 recallZ-0590-2013 · DePuy Orthopaedics, Inc. · initiated Dec 3, 2012

Class II Terminated All recalls posted in December 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

FDA root cause category: Device Design

What the company is doing

Deputh Orthopaedics sent an Urgent Information - Medical Device Recall Notice dated December , 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The notification letter instructed consignees to please cease further distribution or use of products immediately. The purpose of the communication was to inform them of this recall and request acknowledgement of receipt of this letter by their facility. Reconciliation forms should be returned to your DePuy Sales Representative or faxed to the following Fax Number: 574-371-4939. For questions regarding this recall call 574-267-8143.

Recall details

Recall number
Z-0590-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 3, 2012
Posted by the FDA
Dec 20, 2012
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
1732 total all product codes
Product code
LPH
FDA submissions
K060581
Distribution
Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR

Product description

SOL SYS R 9IN CALC 2.25/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Code information (models, lots, serial numbers)

Product Code: 157226195 and Lot codes: B2LAC1000, BN1K6A000, CE3FH1000, CL5PJ1000, X5GB61000, X5GB6A000, X5HCV1000, X5PGT1000, X5PGV1000, X5PGW1000, YF4A81000, YF5DP1000, YG7D61000, and Z41AV1000.

Other devices under the same FDA submission

10 devices cleared under K060581 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 63545).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page