LPS Diaphyseal Sleeves recallZ-0830-2013 · DePuy Orthopaedics, Inc. · initiated Jan 4, 2013
Class I Terminated All recalls posted in February 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.
FDA root cause category: Device Design
What the company is doing
DePuy Othopaedics sent an Urgent Information Medical Device Recall Notice letter dated January 4, 2013 to all affected customers. The letter identified the affected products, reason for recall, clinical implications and actions to be taken. The letter instructed customers to check inventory, discontinue use, cease further distribution and return all inventory on hand for credit. Customers were requested to complete and return the Reconciliation Form. For product-related questions contact your local DePuy Othopaedics Sales Representative, clinical questions contact DePuy's Scientific Information Office at 1-888-554-2482 and recall notice information call 574-372-7333.
Recall details
- Recall number
- Z-0830-2013
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jan 4, 2013
- Posted by the FDA
- Feb 15, 2013
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 162 total units (US) and 102 units total (OUS)
- Product code
- MBH
- FDA submissions
- K071417
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South Africa and Canada including Veteran Administration hospitals.
Product description
LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects.
Code information (models, lots, serial numbers)
Part number: 198720024 and Lot numbers: 113812, 123509, 140578, 156210, 171659, 171660, 171661, 171662, 171663, 123509A, B34GE1000, B34GEA000, B4EGR1000, B4EGRA000, B4EGRB000, B4XFL1000, B4XFLA000, B5SH11000, B5SH1A000, B5SH1B, C74J11000, D14KG1000, E4RF21000, EN8CY1000, EN8CYA000, FC1MB1, and FC1MBA.
Other devices under the same FDA submission
3 devices cleared under K071417 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LPS 1987-06-005Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER51 reports
- LPS 1987-06-010Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- LPS 1987-06-000Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63783).
- Z-0828-2013 · TerminatedLPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct sever…
- Z-0829-2013 · TerminatedLPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct sever…
- Z-0831-2013 · TerminatedLPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct sever…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page