Elekta Synergy XVI. Radiation Therapy Digital Imager.… recallZ-0931-2013 · Elekta, Inc. · initiated Dec 19, 2012
Class II Terminated All recalls posted in March 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
FDA root cause category: Software design
What the company is doing
Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.
Recall details
- Recall number
- Z-0931-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 19, 2012
- Posted by the FDA
- Mar 10, 2013
- Recalling firm
- Elekta, Inc.
- Quantity
- 164
- Product code
- IYE
- FDA submissions
- K051932
- Distribution
- Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
Product description
Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
Other devices under the same FDA submission
1 device cleared under K051932 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 1525717Elekta Limited · Accelerator, linear, medical1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page