Giraffe OmniBed and Giraffe Incubator p/n 6600-0723-700… recallZ-1051-2013 · GE Healthcare, LLC · initiated Jan 8, 2013

Class I Terminated All recalls posted in April 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.

FDA root cause category: Process control

What the company is doing

GE Healthcare sent an "Urgent Medical Device Correction" letter, dated January 8, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were given the safety issue and instructions for continued use of the Giraffe OmniBed and Incubators, as well as given the affected product details. GE Healthcare will correct all affected Giraffe Incubators and Giraffe OmniBeds at no cost, and will contact customers to arrange for this correction. If customers have any questions or concerns regarding this notification, please call: United States: 1-800-345-2700. For other countries, please contact your local GE Healthcare Service Representative.

Recall details

Recall number
Z-1051-2013
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jan 8, 2013
Posted by the FDA
Apr 5, 2013
Recalling firm
GE Healthcare, LLC
Quantity
679 units
Product code
FMZ
FDA submissions
K101778, K101788
Distribution
Worldwide Distribution -- USA in the states of AR ,CA, CO, DC, FL, GA, HI, ID, IL, IN, KS, MA, MD, MI, MO, MS, NC, NE, NJ, NY, NV, OH, PA, TN, TX, VA, WA, WI, and WV, and the countries of Arab Emirates, Albania, Australia, Belgium, Brazil, Canada, Chile, China, Croatia, Czech., Denmark, Ecuador, Estonia, Finland, France, Germany, Hungary, Italy ,Korea, Kurdistan, Lebanon, Macedonia, Malaysia, Mexico, Norway, Oman, Poland, Russia, San Salvador, Saudi Arabia, Singapore, South Africa, Sweden, Turkey, Uruguay, Venezuela, and United Kingdom.

Product description

Giraffe OmniBed and Giraffe Incubator p/n 6600-0723-700 (rev A and B) labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. GE Healthcare Giraffe Omnibed is a combination of an infant incubator and an infant warmer.

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

OmniBed Serial Numbers: HDGR52546, HDGR52736, HDGR52802, HDGR52861, HDGR52910, HDGR52958, HDGR53007, HDGR53058, HDGR53110, HDGR53185, HDGR52547, HDGR52737, HDGR52803, HDGR52862, HDGR52911, HDGR52959, HDGR53008, HDGR53059, HDGR53112, HDGR53186, HDGR52558, HDGR52739, HDGR52804, HDGR52863, HDGR52912, HDGR52960, HDGR53009, HDGR53060, HDGR53113, HDGR53187, HDGR52609, HDGR52740, HDGR52805, HDGR52864, HDGR52913, HDGR52961, HDGR53010, HDGR53061, HDGR53114, HDGR53188, HDGR52610, HDGR52741, HDGR52806, HDGR52865, HDGR52914, HDGR52962, HDGR53011, HDGR53062, HDGR53115, HDGR53189, HDGR52613, HDGR52742, HDGR52813, HDGR52866, HDGR52915, HDGR52963, HDGR53012, HDGR53063, HDGR53116, HDGR53190, HDGR52620, HDGR52743, HDGR52814, HDGR52867, HDGR52916, HDGR52964, HDGR53013, HDGR53064, HDGR53117, HDGR53191, HDGR52621, HDGR52744, HDGR52815, HDGR52868, HDGR52917, HDGR52965, HDGR53014, HDGR53065, HDGR53119, HDGR53192, HDGR52622, HDGR52745, HDGR52816, HDGR52869, HDGR52918, HDGR52966, HDGR53015, HDGR53066, HDGR53120, HDGR53193, HDGR52623, HDGR52746, HDGR52817, HDGR52870, HDGR52919, HDGR52967, HDGR53016, HDGR53067, HDGR53125, HDGR53194, HDGR52624, HDGR52747, HDGR52818, HDGR52871, HDGR52920, HDGR52968, HDGR53018, HDGR53068, HDGR53126, HDGR53195, HDGR52626, HDGR52748, HDGR52820, HDGR52872, HDGR52921, HDGR52969, HDGR53019, HDGR53069, HDGR53127, HDGR53196, HDGR52657, HDGR52749, HDGR52821, HDGR52873, HDGR52922, HDGR52970, HDGR53021, HDGR53070, HDGR53128, HDGR53197, HDGR52658, HDGR52750, HDGR52822, HDGR52874, HDGR52923, HDGR52971, HDGR53022, HDGR53071, HDGR53129, HDGR53213, HDGR52664, HDGR52751, HDGR52823, HDGR52875, HDGR52924, HDGR52972, HDGR53023, HDGR53072, HDGR53134, HDGR53214, HDGR52665, HDGR52752, HDGR52824, HDGR52876, HDGR52925, HDGR52973, HDGR53024, HDGR53073, HDGR53135, HDGR53215, HDGR52668, HDGR52753, HDGR52825, HDGR52877, HDGR52926, HDGR52974, HDGR53025, HDGR53074, HDGR53136, HDHR53216, HDGR52669, HDGR52754, HDGR52827, HDGR52878, HDGR52927, HDGR52975, HDGR53026, HDGR53075, HDGR53137, HDGR53217, HDGR52670, HDGR52755, HDGR52828, HDGR52879, HDGR52928, HDGR52976, HDGR53027, HDGR53076, HDGR53138, HDGR53219, HDGR52671, HDGR52756, HDGR52829, HDGR52880, HDGR52929, HDGR52977, HDGR53028, HDGR53077, HDGR53139, HDGR53220, HDGR52672, HDGR52757, HDGR52830, HDGR52881, HDGR52930, HDGR52978, HDGR53029, HDGR53078, HDGR53140, HDGR53221, HDGR52673, HDGR52758, HDGR52831, HDGR52882, HDGR52931, HDGR52979, HDGR53030, HDGR53079, HDGR53141, HDGR53222, HDGR52674, HDGR52759, HDGR52832, HDGR52883, HDGR52932, HDGR52980, HDGR53031, HDGR53080, HDGR53142, HDGR53223, HDGR52675, HDGR52760, HDGR52833, HDGR52884, HDGR52933, HDGR52981, HDGR53032, HDGR53081, HDGR53145, HDGR53224, HDGR52678, HDGR52761, HDGR52835, HDGR52885, HDGR52934, HDGR52982, HDGR53033, HDGR53085, HDGR53146, HDGR53225, HDGR52682, HDGR52762, HDGR52836, HDGR52886, HDGR52935, HDGR52983, HDGR53034, HDGR53086, HDGR53147, HDGR53226, HDGR52683, HDGR52763, HDGR52839, HDGR52887, HDGR52936, HDGR52984, HDGR53035, HDGR53087, HDGR53152, HDGR53231, HDGR52684, HDGR52764, HDGR52840, HDGR52888, HDGR52937, HDGR52985, HDGR53036, HDGR53088, HDGR53153, HDGR53232, HDGR52685, HDGR52765, HDGR52841, HDGR52890, HDGR52938, HDGR52986, HDGR53037, HDGR53089, HDGR53154, HDGR53235, HDGR52686, HDGR52768, HDGR52842, HDGR52891, HDGR52939, HDGR52987, HDGR53038, HDGR53090, HDGR53155, HDGR53241, HDGR52687, HDGR52769, HDGR52843, HDGR52892, HDGR52940, HDGR52988, HDGR53039, HDGR53091, HDGR53156, HDGR53242, HDGR52700, HDGR52770, HDGR52844, HDGR52893, HDGR52941, HDGR52990, HDGR53040, HDGR53092, HDGR53157, HDGR53251, HDGR52709, HDGR52771, HDGR52845, HDGR52894, HDGR52942, HDGR52991, HDGR53041, HDGR53093, HDGR53158, HDGR52714, HDGR52775, HDGR52846, HDGR52895, HDGR52943, HDGR52992, HDGR53042, HDGR53094, HDGR53159, HDGR52718, HDGR52776, HDGR52847, HDGR52896, HDGR52944, HDGR52993, HDGR53043, HDGR53095, HDGR53160, HDGR52720, HDGR52777, HDGR52848, HDGR52897, HDGR52945, HDGR52994, HDGR53044, HDGR53096, HDGR53161, HDGR52721, HDGR52778, HDGR52849, HDGR52898, HDGR52946, HDGR52995, HDGR53045, HDGR53097, HDGR53162, HDGR52723, HDGR52779, HDGR52850, HDGR52899, HDGR52947, HDGR52996, HDGR53047, HDGR53098, HDGR53163, HDGR52726, HDGR52780, HDGR52851, HDGR52900, HDGR52948, HDGR52997, HDGR53048, HDGR53099, HDGR53164, HDGR52727, HDGR52781, HDGR52852, HDGR52901, HDGR52949, HDGR52998, HDHR53049, HDGR53100, HDGR53165, HDGR52728, HDGR52782, HDGR52853, HDGR52902, HDHR52950, HDGR52999, HDGR53050, HDGR53102, HDGR53166, HDGR52729, HDGR52789, HDGR52854, HDGR52903, HDGR52951, HDGR53000, HDGR53051, HDGR53103, HDGR53167, HDGR52730, HDGR52790, HDGR52855, HDGR52904, HDGR52952, HDGR53001, HDGR53052, HDGR53104, HDGR53168, HDGR52731, HDGR52791, HDGR52856, HDGR52905, HDGR52953, HDGR53002, HDGR53053, HDGR53105, HDGR53169, HDGR52732, HDGR52792, HDGR52857, HDGR52906, HDGR52954, HDGR53003, HDGR53054, HDGR53106, HDGR53171, HDGR52733, HDGR52793, HDGR52858, HDGR52907, HDGR52955, HDGR53004, HDGR53055, HDGR53107, HDGR53175, HDGR52734, HDGR52794, HDGR52859, HDGR52908, HDGR52956, HDGR53005, HDGR53056, HDGR53108, HDGR53176, HDGR52735, HDGR52799, HDGR52860, HDGR52909, HDGR52957, HDGR53006, HDGR53057, HDGR53109, and HDGR53181. ***Incubator Serial Numbers: HDHR52549, HDHR52708, HDHR52774, HDHR52881, HDHR53029, HDHR52551, HDHR52709, HDHR52830, HDHR52882, HDHR53030, HDHR52552, HDHR52710, HDHR52831, HDHR52883, HDHR53031, HDHR52553, HDHR52711, HDHR52832, HDHR52884, HDHR53032, HDHR52554, HDHR52720, HDHR52833, HDHR52885, HDHR53039, HDHR52555, HDHR52721, HDHR52834, HDHR52894, HDHR53040, HDHR52556, HDHR52722, HDHR52835, HDHR52900, HDHR53041, HDHR52557, HDHR52723, HDHR52836, HDHR52901, HDHR53042, HDHR52559, HDHR52724, HDHR52837, HDHR52902, HDHR53043, HDHR52561, HDHR52725, HDHR52838, HDHR52903, HDHR53044, HDHR52562, HDHR52726, HDHR52839, HDHR52904, HDHR53045, HDHR52563, HDHR52727, HDHR52840, HDHR52905, HDHR53046, HDHR52564, HDHR52728, HDHR52841, HDHR52906, HDHR53047, HDHR52565, HDHR52729, HDHR52842, HDHR52907, HDHR53048, HDHR52567, HDHR52730, HDHR52843, HDHR52908, HDHR53049, HDHR52568, HDHR52731, HDHR52844, HDHR52909, HDHR53030, HDHR52569, HDHR52732, HDHR52845, HDHR52951, HDHR52910, HDHR52572, HDHR52733, HDHR52847, HDHR52911, HDHR53052, HDHR52573, HDHR52734, HDHR52848, HDHR52912, HDHR53053, HDHR52575, HDHR52735, HDHR52849, HDHR52913, HDHR53054, HDHR52576, HDHR52736, HDHR52850, HDHR52914, HDHR53055, HDHR52577, HDHR52737, HDHR52851, HDHR52915, HDHR53059, HDHR52578, HDHR52738, HDHR52852, HDHR52926, HDHR53062, HDHR52579, HDHR52739, HDHR52853, HDHR52927, HDHR52580, HDHR52740, HDHR52854, HDHR52928, HDHR52581, HDHR52741, HDHR52855, HDHR52929, HDHR52588, HDHR52742, HDHR52857, HDHR52971, HDHR52589, HDHR52743, HDHR52858, HDHR52972, HDHR52650, HDHR52744, HDHR52859, HDHR52974, HDHR52653, HDHR52745, HDHR52860, HDHR52975, HDHR52654, HDHR52746, HDHR52861, HDHR52976, HDHR52689, HDHR52747, HDHR52862, HDHR52977, HDHR52690, HDHR52748, HDHR52863, HDHR52978, HDHR52691, HDHR52749, HDHR52864, HDHR52979, HDHR52692, HDHR52750, HDHR52865, HDHR52981, HDHR52693, HDHR52751, HDHR52866, HDHR52982, HDHR52694, HDHR52752, HDHR52867, HDHR52983, HDHR52695, HDHR52753, HDHR52868, HDHR52985, HDHR52696, HDHR52754, HDHR52869, HDHR52986, HDHR52697, HDHR52759, HDHR52870, HDHR52987, HDHR52698, HDHR52760, HDHR52871, HDHR52988, HDHR52699, HDHR52767, HDHR52872, HDHR52989, HDHR52700, HDHR52768, HDHR52873, HDHR52990, HDHR52701, HDHR52769, HDHR52874, HDHR52991, HDHR52702, HDHR52770, HDHR52875, HDHR52992, HDHR52703, HDHR52771, HDHR52876, HDHR52993, HDHR52704, HDHR52772, HDHR52877, HDHR52994, HDHR52707, HDHR52773, HDHR52880, and HDHR52995.

Other devices under the same FDA submission

1 device cleared under K101778, K101788 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.