Medical device recalls: April 2013
The FDA posted 162 medical device recall records in April 2013, from 120 recall events by 78 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 11 events, 11 records, 1 Class I
- Siemens Healthcare Diagnostics, Inc. · 4 events, 6 records
- Toshiba American Medical Systems Inc · 4 events, 4 records
- Siemens Medical Solutions USA, Inc · 3 events, 11 records
- Hospira Inc. · 3 events, 6 records, 4 Class I
- Ortho-Clinical Diagnostics · 3 events, 5 records
- Edwards Lifesciences, LLC · 3 events, 4 records
- Advanced Sterilization Products · 3 events, 3 records
FDA root cause
- Device Design · 28
- Nonconforming Material/Component · 21
- Software design · 20
- Process control · 14
- Software design (manufacturing process) · 9
- Labeling False and Misleading · 9
- Process design · 7
- No Marketing Application · 7
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1169-2013 · Hospira Inc. · posted Apr 30, 2013 · Terminated
Pump shutting off during use without warning.
- Z-1170-2013 · Hospira Inc. · posted Apr 30, 2013 · Terminated
Pump shutting off during use without warning.
- GemStar Pump - Battery ↗Class IZ-1159-2013 · Hospira Inc. · posted Apr 26, 2013 · Terminated
Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
- GemStar Pump - Battery ↗Class IZ-1160-2013 · Hospira Inc. · posted Apr 26, 2013 · Terminated
Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
- Z-1171-2013 · Newport Medical Instruments Inc · posted Apr 26, 2013 · Terminated
Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup…
- Z-1177-2013 · Maquet Cardiovascular Us Sales, Llc · posted Apr 26, 2013 · Terminated
When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.
- Z-1098-2013 · Carefusion Corporation · posted Apr 15, 2013 · Terminated
The recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.
- ShapeMatch Cutting Guides ↗Class IZ-1052-2013 · Stryker Howmedica Osteonics Corp. · posted Apr 12, 2013 · Terminated
With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing ShapeMatch Cutting Guides.
- Z-1074-2013 · Lifescan Inc · posted Apr 12, 2013 · Terminated
The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
- Z-1058-2013 · Engineered Medical Systems,Inc · posted Apr 10, 2013 · Terminated
The mask inside the package has been identified as "Neonatal" in size and not "Infant" as the labeling indicates.
- Z-1051-2013 · GE Healthcare, LLC · posted Apr 5, 2013 · Terminated
Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.
- Z-0993-2013 · Animas Corporation · posted Apr 3, 2013 · Terminated
Animas has identified a component issue affecting a small supply of the Animas(R) 2020 insulin pumps. The component issue may trigger the pumps to sound a false alarm or warning related to one of the following: - Loss of prime; -Occlusion; or -No…
Class II recalls (133)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1184-2013 · Covidien LLC · posted Apr 29, 2013 · Terminated
Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient
- Z-1185-2013 · Covidien LLC · posted Apr 29, 2013 · Terminated
Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient
- Z-1186-2013 · Discus Dental LLC · posted Apr 29, 2013 · Terminated
The firm is initiating this recall of the SL3 Soft Tissue Laser equipped with a fiber optic Unifiber system which contains a fragile glass filament because the fiber broke near the hand piece.
- Rad-8 Pulse Oximeters ↗Class IIZ-1187-2013 · Masimo Corporation · posted Apr 29, 2013 · Terminated
Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button.
- Z-1188-2013 · GE Healthcare, LLC · posted Apr 29, 2013 · Terminated
Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within…
- Z-1189-2013 · Philips Healthcare Inc. · posted Apr 29, 2013 · Terminated
Device may fail to deliver defibrillation therapy in either Manual Defib or AED mode.
- Z-1180-2013 · Advanced Sterilization Products · posted Apr 26, 2013 · Terminated
Advanced Sterilization Products is recalling certain lots of STERRAD CYCLESURE 24 Biological Indicator product because it may not have adequate data to support the entire duration of the labeled shelf-life.
- Z-1181-2013 · Vital Images, Inc. · posted Apr 26, 2013 · Terminated
Vital Images, Inc. (Vital Images) received one report about potential user confusion when viewing a brain perfusion scan on Vitrea CT Brain Perfusion 2D when the scan was performed using a wide coverage shuttle acquisition technique. This may result in…
- Z-1182-2013 · Siemens Medical Solutions USA, Inc · posted Apr 26, 2013 · Terminated
Software update to fix multiple safety related issues.
- Z-1183-2013 · B. Braun Medical, Inc. · posted Apr 26, 2013 · Terminated
There is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient).
- Z-1172-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Apr 25, 2013 · Terminated
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue…
- ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113.Class IIZ-1173-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Apr 25, 2013 · Terminated
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue…
- Z-1174-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Apr 25, 2013 · Terminated
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue…
- Z-1175-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Apr 25, 2013 · Terminated
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue…
- Z-1176-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Apr 25, 2013 · Terminated
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue…
- Z-1178-2013 · Alere San Diego, Inc. · posted Apr 25, 2013 · Terminated
The Alere Cholestech LDX hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the…
- Z-1167-2013 · Beckman Coulter Inc. · posted Apr 24, 2013 · Terminated
Beckman Coulter, Inc. initiated this recall because there is a software error that has the potential to cause the reporting of erroneous patient results because of misidentification by the AU5800 system.
- Z-1168-2013 · GE Healthcare, LLC · posted Apr 24, 2013 · Terminated
GE Healthcare has recently become aware of an incident which caused a patient injury during an unload following an exam on a Ventri scanner. During a patient unload following a scan, the patients fingers may be pinched between the moving cradle edge and…
- Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 ↗Class IIZ-1163-2013 · Varian Medical Systems, Inc. Oncology Systems · posted Apr 23, 2013 · Terminated
Varian has identified an anomaly with the RPM Respiratory Gating System version I.7.5 with 3D Option. While operating with the 6-dot marker block, this version of RPM Gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions.
- Z-1164-2013 · GE Healthcare, LLC · posted Apr 23, 2013 · Terminated
GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts…
- Z-1165-2013 · Philips And Neusoft Medical Systems Co., Ltd. · posted Apr 23, 2013 · Terminated
User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
- Z-1166-2013 · Philips And Neusoft Medical Systems Co., Ltd. · posted Apr 23, 2013 · Terminated
User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
- Z-1080-2013 · Stellate Systems · posted Apr 22, 2013 · Terminated
The firm is initiating this recall because the implied or stated indications in their labeling associated with Harmonie software have not yet been cleared by the FDA for sale in the USA.
- Z-1156-2013 · Orthopedic Alliance LLC · posted Apr 22, 2013 · Terminated
Orthopedic Alliance is recalling Blue & Gold Implants and Instruments because the Blue & Gold system is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in…
- Z-1157-2013 · Spinal Solutions, LLC · posted Apr 22, 2013 · Terminated
Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could…
- Z-1158-2013 · Teleflex Medical · posted Apr 22, 2013 · Terminated
Sterile packaging may be compromised.
- Z-1155-2013 · Accuray Incorporated · posted Apr 20, 2013 · Terminated
A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.
- Z-1152-2013 · Ortho-Clinical Diagnostics · posted Apr 19, 2013 · Terminated
Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. OCD's investigation…
- Z-1153-2013 · Nova Biomedical Corporation · posted Apr 19, 2013 · Terminated
Glucose test strips reports no result message when tested
- Z-1154-2013 · Nova Biomedical Corporation · posted Apr 19, 2013 · Terminated
Glucose test strips reports no result message when tested
- Z-1131-2013 · Heritage Labs International LLC · posted Apr 18, 2013 · Terminated
The product has a Lithium Heparin contaminant which could adversely affect coagulation results potentially causing erroneous lab results.
- Z-1140-2013 · Nipro Diagnostics, Inc. · posted Apr 18, 2013 · Terminated
Meter kit outer boxes containing 10 ct. test strips state incorrect storage temperature range.
- Z-1141-2013 · Horiba Instruments, Inc dba Horiba Medical · posted Apr 18, 2013 · Terminated
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have…
- Z-1142-2013 · GE Healthcare, LLC · posted Apr 18, 2013 · Terminated
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor,…
- Elana Catheter Sizer 2.0*** ↗Class IIZ-1143-2013 · Elana, Inc. · posted Apr 18, 2013 · Terminated
Distribution of an unapproved device into interstate commerce.
- Z-1144-2013 · Vital Images, Inc. · posted Apr 18, 2013 · Terminated
Vital Images has made a decision to perform a field correction to the Vitrea CT Multi-Chamber Cardiac Functional Analysis application. Vital Images has identified a defect in which numerical values may be incorrect in the Results Table under the Cardiac…
- ACIST Medical Systems Inc. ↗Class IIZ-1145-2013 · Acist Medical Systems · posted Apr 18, 2013 · Terminated
ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the…
- Z-1146-2013 · Edwards Lifesciences, LLC · posted Apr 18, 2013 · Terminated
The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.
- Z-1147-2013 · Transonic Systems Inc · posted Apr 18, 2013 · Terminated
Transonic Systems Inc. have received complaints of blood leakage involving the AV Loop Kit.
- Z-1149-2013 · Maquet Cardiovascular, LLC · posted Apr 18, 2013 · Terminated
It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the…
- Z-1150-2013 · Advanced Sterilization Products · posted Apr 18, 2013 · Terminated
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold…
- Z-1151-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 18, 2013 · Terminated
Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal…
- Z-1003-2013 · Philips Healthcare Inc. · posted Apr 17, 2013 · Terminated
Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.
- DeGotzen XRay unit XGenus ↗Class IIZ-1020-2013 · DeGotzen · posted Apr 17, 2013 · Terminated
It was discovered at the WEAC Labs method sample did not meet requirements for labeling and certification
- The Oxford Partial Knee ↗Class IIZ-1127-2013 · Synvasive Technology Inc · posted Apr 17, 2013 · Terminated
Biomet part # 506076, lot 928182 was received from Synvasive Technology containing the incorrect blade.
- Z-1128-2013 · Sr Instruments Inc · posted Apr 17, 2013 · Terminated
SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in…
- LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. ↗Class IIZ-1129-2013 · Lifescan Inc · posted Apr 17, 2013 · Terminated
When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
- Z-1130-2013 · Lifescan Inc · posted Apr 17, 2013 · Terminated
When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
- Z-1132-2013 · Smith & Nephew Inc · posted Apr 17, 2013 · Terminated
One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter).
- Z-1133-2013 · GE Healthcare, LLC · posted Apr 17, 2013 · Terminated
GE Healthcare conducted a recall on the Muse v7 because incorrect medications were displayed on the MUSE, which could result in incorrect treatment.
- Z-1134-2013 · Edwards Lifesciences, LLC · posted Apr 17, 2013 · Terminated
Holes in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
- Z-1135-2013 · Edwards Lifesciences, LLC · posted Apr 17, 2013 · Terminated
Holes in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
- Z-1136-2013 · Endogastric Solutions Inc · posted Apr 17, 2013 · Terminated
EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.
- GE Healthcare, AW Server.Class IIZ-1137-2013 · GE Healthcare, LLC · posted Apr 17, 2013 · Terminated
GE Healthcare has recently become aware of a potential safety issue with respect to the Results Viewer in the AW Server. After starting the Results Viewer with the selected series of a patient, the pixel and certain annotation data from a different…
- OsteoSelect DBM Putty ↗Class IIZ-1138-2013 · Bacterin International, Inc. · posted Apr 17, 2013 · Terminated
The firm's retesting procedure was not validated.
- Z-1139-2013 · Alere San Diego, Inc. · posted Apr 17, 2013 · Terminated
The recall was initiated because Alere San Diego has an update for the Alere Cholestech LDX ALT AST cassette, and it has the potential for humidity changes to impact results for certain analytes on the Alere Cholestech LDX System.
- Z-1125-2013 · Instrumentation Industries Inc · posted Apr 16, 2013 · Terminated
Firm initiated an update to Directions for Use necessitating replacement of previously distributed Directions for Use.
- Z-1126-2013 · Accuray Incorporated · posted Apr 16, 2013 · Terminated
A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans…
- Z-1112-2013 · Carefusion Corporation · posted Apr 15, 2013 · Terminated
The recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic medical record…
- Z-1113-2013 · Roche Diagnostics Operations, Inc. · posted Apr 15, 2013 · Terminated
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.
- Z-1114-2013 · Roche Diagnostics Operations, Inc. · posted Apr 15, 2013 · Terminated
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality…
- Z-1115-2013 · Synvasive Technology Inc · posted Apr 15, 2013 · Terminated
Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors include: (1) The pouch says oscillating saw blade when it actually contains a reciprocating saw blade; (2) The box bar code date does not match the human…
- Z-1116-2013 · Hospira Inc. · posted Apr 15, 2013 · Terminated
During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
- Z-1117-2013 · Hospira Inc. · posted Apr 15, 2013 · Terminated
During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
- Z-1118-2013 · Abbott Laboratories, Inc · posted Apr 15, 2013 · Terminated
Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal…
- Z-1119-2013 · Siemens Healthcare Diagnostics, Inc · posted Apr 15, 2013 · Terminated
Negative bias in patient samples for Vitamin D
- Z-1120-2013 · Siemens Healthcare Diagnostics, Inc · posted Apr 15, 2013 · Terminated
Negative bias in patient samples for Vitamin D
- Z-1121-2013 · Siemens Healthcare Diagnostics, Inc · posted Apr 15, 2013 · Terminated
Negative bias in patient samples for Vitamin D
- Z-1122-2013 · Siemens Healthcare Diagnostics, Inc · posted Apr 15, 2013 · Terminated
Negative bias in patient samples for Vitamin D
- Z-1123-2013 · Advanced Sterilization Products · posted Apr 15, 2013 · Terminated
Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.
- REF 32-422760 ↗Class IIZ-1124-2013 · Biomet U.K., Ltd. · posted Apr 15, 2013 · Terminated
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
- Z-1111-2013 · Edwards Lifesciences, LLC · posted Apr 14, 2013 · Terminated
Two unsealed pouches were found at an Edwards' distribution location in Japan. Incorrectly sealed pouches may result in a sterility barrier breach.
- Z-1106-2013 · Ortho-Clinical Diagnostics · posted Apr 13, 2013 · Terminated
Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its…
- Z-1107-2013 · Ortho-Clinical Diagnostics · posted Apr 13, 2013 · Terminated
Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its…
- Z-1108-2013 · Ortho-Clinical Diagnostics · posted Apr 13, 2013 · Terminated
Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its…
- Z-1109-2013 · Intuitive Surgical, Inc. · posted Apr 13, 2013 · Terminated
Revised instructions for users of the da Vinci S, Si and Si-e systems with additional information regarding sterile adapter engagement and instrument installation.
- Z-1110-2013 · Intuitive Surgical, Inc. · posted Apr 13, 2013 · Terminated
Revised instructions for users of the da Vinci S, Si and Si-e systems with additional information regarding sterile adapter engagement and instrument installation.
- Z-1102-2013 · Philips Medical Systems (Cleveland) Inc · posted Apr 12, 2013 · Terminated
Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after…
- Z-1103-2013 · Integra LifeSciences Corp. · posted Apr 12, 2013 · Terminated
Depending on the system configuration, a software error message in versions 5.0.1 and 5.0.2. occurs if a beam plan is transmitted from XKnife using DICOM-RT.
- Z-1104-2013 · Toshiba American Medical Systems Inc · posted Apr 12, 2013 · Terminated
Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been…
- Z-1105-2013 · Siemens Medical Solutions USA, Inc · posted Apr 12, 2013 · Terminated
Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or…
- Z-1096-2013 · GE Healthcare, LLC · posted Apr 11, 2013 · Terminated
GE Healthcare has become aware of a potential safety issue due to the S4-10-D Ultrasound Probe used in conjunction with the LOGIQ S8 and LOGIQ S7 Ultrasound systems. It was also discovered that this issue also leads to a specification issue of the acoustic…
- Z-1097-2013 · Siemens Medical Solutions USA, Inc · posted Apr 11, 2013 · Terminated
Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is…
- Torrent" irrigation tubing. ↗Class IIZ-1099-2013 · US Endoscopy Group Inc · posted Apr 11, 2013 · Terminated
The firm was notified by their customers that there was a water leakage from the white tubing connector.
- Z-1100-2013 · GE Healthcare, LLC · posted Apr 11, 2013 · Terminated
GE Healthcare has recently become aware of a potential issue due to image flip in frequency encoding direction associated with a specific workflow of your Discovery MR750w system. Image flip issue in frequency encoding direction (including S/I, A/P and…
- MRT-1504/A4 - Titan X MRI SystemClass IIZ-1101-2013 · Toshiba American Medical Systems Inc · posted Apr 11, 2013 · Terminated
It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank.…
- Z-1076-2013 · Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical · posted Apr 10, 2013 · Terminated
Kerr Corporation is initiating this recall because raw material used to produce the Maxcem Elite is causing the material to prematurely polymerize in the syringe. The product could be difficult to extrude or demonstrate a shorter work-time as the material…
- Z-1077-2013 · Siemens Medical Solutions USA, Inc · posted Apr 10, 2013 · Terminated
A potential safety risk with the MOSAIQ Oncology Information System that has been identified by the manufacturer IMPAC Medical Systems, Inc.
- Z-1078-2013 · GE Healthcare, LLC · posted Apr 10, 2013 · Terminated
Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.
- Z-1079-2013 · Thermo Fisher Scientific (Asheville) LLC · posted Apr 10, 2013 · Terminated
Incubators manufactured prior to mid 2009 have coated metal hinges that have reportedly failed in the field.
- Z-1081-2013 · Siemens Medical Solutions USA, Inc. · posted Apr 10, 2013 · Terminated
When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to…
- Z-1082-2013 · Physio Control, Inc. · posted Apr 10, 2013 · Terminated
LIFEPAK CR Plus or LIFEPAK Express Automated External Defibrillators (AEDs) are recalled due to the potential to experience an early internal battery depletion issue.
- Z-1083-2013 · GE Healthcare, LLC · posted Apr 10, 2013 · Terminated
GE Healthcare has become aware of an issue associated with the base casting for the articulated arm of your VMX and VMX Plus Mobile X-ray systems that may impact Operator or Patient safety.
- Z-1084-2013 · Ormco/Sybronendo · posted Apr 10, 2013 · Terminated
Ormco Corporation is voluntarily recalling one (1) lot of Bracket Buccal Tubes as a result of a packaging error. More specifically, the Bracket Buccal Tubes product package correctly identifies the product as gingival (0) offset brackets, but actually…
- Z-1085-2013 · Medtronic MiniMed · posted Apr 10, 2013 · Terminated
Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the…
- Z-1071-2013 · J T Posey Company · posted Apr 9, 2013 · Terminated
The firm voluntarily recalled Biothane Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to become disconnected from the connecting strap and allow patients…
- Precise Treatment Table. ↗Class IIZ-1072-2013 · Elekta, Inc. · posted Apr 9, 2013 · Terminated
The Column rotation can accidentally be moved to a different position which in the locked position.
- Z-1073-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 9, 2013 · Terminated
Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.
- TSX-IOI: AQUILION 32/64 ↗Class IIZ-1067-2013 · Toshiba American Medical Systems Inc · posted Apr 8, 2013 · Terminated
The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.
- Z-1068-2013 · Toshiba American Medical Systems Inc · posted Apr 8, 2013 · Terminated
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
- Z-1069-2013 · AGFA Corp. · posted Apr 8, 2013 · Terminated
Software design error in IMPAX CV 7.8 SU3
- Z-1070-2013 · AGFA Corp. · posted Apr 8, 2013 · Terminated
Baseline pulmonary capillary wedge (PCW) pressure values from IMPAX CV Reporting Cardiac Catheterization module were not correctly transferring into the hospital's electronic health record (EHR) system.
- Z-1066-2013 · Stelkast Co · posted Apr 7, 2013 · Terminated
Firm was made aware of incidents relating to the difficulty of properly engaging and securing the Cross-Over 36mm liner into the Cross-Over shell.
- Z-1064-2013 · Arrow International Inc · posted Apr 6, 2013 · Terminated
Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
- Z-1065-2013 · Arrow International Inc · posted Apr 6, 2013 · Terminated
Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
- Z-1061-2013 · Stryker Howmedica Osteonics Corp. · posted Apr 5, 2013 · Terminated
Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly.
- Z-1062-2013 · Synthes USA HQ, Inc. · posted Apr 5, 2013 · Terminated
Synthes is initiating a voluntary recall of the Flexible Grip (Part number 355.28) which is part of the Universal Nail System, due to the potential lack of the ability to thoroughly clean the instrument.
- Z-1063-2013 · Covidien LP · posted Apr 5, 2013 · Terminated
Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
- Z-1054-2013 · LeMaitre Vascular, Inc. · posted Apr 4, 2013 · Terminated
Devices were not sealed correctly during the manufacturing process, and the sterility of these products have been compromised.
- Z-1055-2013 · LeMaitre Vascular, Inc. · posted Apr 4, 2013 · Terminated
Devices were not sealed correctly during the manufacturing process, and the sterility of these products have been compromised.
- Z-1057-2013 · Arrow International Inc · posted Apr 4, 2013 · Terminated
The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the…
- Z-1059-2013 · Polymedco, Inc · posted Apr 4, 2013 · Terminated
The test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life. There may be a possibility of obtaining false negative results when…
- Z-1060-2013 · Polymedco, Inc · posted Apr 4, 2013 · Terminated
The test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life. There may be a possibility of obtaining false negative results when…
- LEEP REDIKIT Product No. 6060 ↗Class IIZ-1053-2013 · Cooper Surgical, Inc. · posted Apr 3, 2013 · Terminated
LEEP RediKit Products (for use in Loop Electrosurgical Excision Procedures) have Xylocaine (Lidocaine) HCL (2%) listed on the tray lid label but this is not included in the contents of the kit, which contained Polocaine (Mepivacaine) HCL (3%) instead…
- Z-1031-2013 · Medtronic Neuromodulation · posted Apr 2, 2013 · Terminated
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off.…
- Z-1036-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1037-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1038-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1039-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1040-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1041-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1042-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1043-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1044-2013 · Siemens Medical Solutions USA, Inc · posted Apr 2, 2013 · Terminated
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several…
- Z-1049-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 2, 2013 · Terminated
Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells that are in close proximity to open…
- Z-1028-2013 · Roche Diagnostics Operations, Inc. · posted Apr 1, 2013 · Terminated
Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.
- Ingenuity TF PET/CT. ↗Class IIZ-1029-2013 · Philips Medical Systems (Cleveland) Inc · posted Apr 1, 2013 · Terminated
Issues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, including inaccurate standard uptake values (SUV). Philips…
- Z-1030-2013 · Integra Burlington MA, Inc. · posted Apr 1, 2013 · Terminated
If the headphone jack is in use during a surgical procedure with the OCS2 and a non-intended voltage (such as static electricity) comes in contact with the outer case of the unit, that voltage could be transmitted to the patient and could be a potential…
- Access Immunoassay System ↗Class IIZ-1032-2013 · Beckman Coulter Inc. · posted Apr 1, 2013 · Terminated
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the…
- Access 2 Immunoassay System ↗Class IIZ-1033-2013 · Beckman Coulter Inc. · posted Apr 1, 2013 · Terminated
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the…
- Z-1034-2013 · Beckman Coulter Inc. · posted Apr 1, 2013 · Terminated
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the…
- Z-1035-2013 · Beckman Coulter Inc. · posted Apr 1, 2013 · Terminated
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the…
- Z-1045-2013 · Zimmer, Inc. · posted Apr 1, 2013 · Terminated
Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of…
Class III recalls (17)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1161-2013 · Ortho-Clinical Diagnostics · posted Apr 23, 2013 · Terminated
Ortho Clinical Diagnostics is recalling certain lots of the Vitros Anti-HBs Reagent packs an inability to obtain a valid calibration.
- Z-1162-2013 · Sirtex Medical Limited · posted Apr 23, 2013 · Terminated
Incorrect expiration date
- IMO-402 Mononucleosis Rapid Test Device (Whole Blood/Serum/Plasma). Catalog #IMO-402/94201 ↗Class IIIZ-1148-2013 · Alere San Diego, Inc. · posted Apr 18, 2013 · Terminated
Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.
- G7 HPLC. ↗Class IIIZ-1075-2013 · Tosoh Bioscience Inc · posted Apr 10, 2013 · Terminated
The Beta-thalessemia Mode of the G7 Analyzer-Beta Thalassemia Mode is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports HbF and HbA2 quantitative percentages in whole blood. The firm received complaints that the…
- Z-1086-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1087-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- LHB, First Aid Type I Individual Kit, NSN: 6545-00-656-1092, antiseptics, Kit number: 2114002. ↗Class IIIZ-1088-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1089-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1090-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1091-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1092-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1093-2013 · Lighthouse For The Blind · posted Apr 10, 2013 · Terminated
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility…
- Z-1056-2013 · Becton Dickinson & Company · posted Apr 4, 2013 · Terminated
The BD Vacutainer Urinalysis Transfer Straw Kit #364991, Lot 1335484 has preprinted polybags which are incorrectly labeled as Catalog #364953. BD advises any Customer who purchased this Kit that the Tube in this Kit is not a C&S tube. If a Customer relies…
- Z-1050-2013 · Horiba Instruments, Inc dba Horiba Medical · posted Apr 3, 2013 · Terminated
Horiba Medical is recalling ABX Micros IM2 Data Management System because it may display incorrect patient test results under unanticipated operating conditions.
- Z-1046-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 1, 2013 · Terminated
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to…
- Z-1047-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 1, 2013 · Terminated
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to…
- Z-1048-2013 · Siemens Healthcare Diagnostics, Inc. · posted Apr 1, 2013 · Terminated
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.