Firebird Spinal Fixation System, Parallel Rod Connector… recallZ-0896-2013 · Orthofix, Inc · initiated Feb 1, 2013

Class II Terminated All recalls posted in February 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

FDA root cause category: Device Design

What the company is doing

Orthofix sent an Urgent Medical Device Recall Notification dated February 1, 2013 to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to identify and remove the affected products from inventory immediately and follow the actions listed in the enclosed Product Return Instructions. For questions or concerns contact Regulatory Affairs at 214-937-2100.

Recall details

Recall number
Z-0896-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 1, 2013
Posted by the FDA
Feb 28, 2013
Recalling firm
Orthofix, Inc
Quantity
565 total (287 PN 52-6800 and 278 PN 52- 6700)
Product code
KWP
FDA submissions
K080407
Distribution
Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.

Product description

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Code information (models, lots, serial numbers)

Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers

Other devices under the same FDA submission

2 devices cleared under K080407 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.