Medtronic DBS Lead Kit for Deep Brain Stimulation, models… recallZ-1190-2013 · Medtronic Neuromodulation · initiated Feb 8, 2013
Class I Terminated All recalls posted in May 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag
FDA root cause category: Process control
What the company is doing
Medtronic sent an Urgent Medical Device Correction letter dated February 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to follow the modified instructions if the lead cap is used during the implant procedure. For questions call Medtronic Neuromodulation Technical Services 1-800-707-0933. UPDATE: Medtronic mailed an Urgent: Medical Device Removal notification to customers on June 07, 2013, who may have unused DBS lead kits that were manufactured prior to the process change to correct the issue. On June 10, 2013 Medtronic representatives began visiting the identified accounts to facilitate removal of any unused product. The Medtronic Rep documented the hospital visits on an Account Specific Customer Confirmation Form and left a copy with the Hospital.
Recall details
- Recall number
- Z-1190-2013
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Feb 8, 2013
- Posted by the FDA
- May 1, 2013
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 30,000
- Product code
- MHY
- FDA submissions
- P960009
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of: Algeria, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam.
Product description
Medtronic DBS Lead Kit for Deep Brain Stimulation, models 3387, 3387S, 3389, 3389S, 3391, 3391S. Sterile and Non-Pyrogenic. Product Usage: The deep brain stimulation system delivers electrical stimulation to selected targets in the brain. DBS Leads are indicated for Dystonia, Essential Tremor (ET), Obsessive-Compulsive Disorder (OCD), Parkinsons Disease (PD), and epilepsy.
Code information (models, lots, serial numbers)
all product is affected as the lead kits are not being removed from the market. Medtronic is providing Additional Instructions for Capping the Lead and Removing the Lead Cap.
Other devices under the same FDA submission
90 devices cleared under P960009 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ACTIVA RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2,839 reports
- Activa® RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR2 deaths1,849 reports
- ACTIVA PC 37601Medtronic, Inc. · implanted subcortical electrical Stimulator (motor disorders)15 deaths1,625 reports
- Activa® 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1,325 reports
- Activa® 37603Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR8 deaths1,093 reports
- Percept™ RC B35300Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 recall3 deaths796 reports
- Activa® 37602Medtronic, Inc. · Deep brain stimulator for obsessive compulsive disorder (OCD)1 recall6 deaths777 reports
- Percept™ PC BrainSense™ B35200Medtronic, Inc. · Deep brain stimulator for obsessive compulsive disorder (OCD)4 recalls1 death547 reports
- Percept™ PC BrainSense™ B35200Medtronic, Inc. · Stimulator, thalamic, epilepsy, implanted1 recall5 deaths543 reports
- Percept™ PC BrainSense™ B35200Medtronic, Inc. · Stimulator, thalamic, epilepsy, implanted1 recall478 reports
- Activa® PC 37601Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS3 deaths521 reports
- Activa® PC 37601Medtronic, Inc. · implanted subcortical electrical Stimulator (motor disorders)1 death502 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page