Lifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 recallZ-0914-2013 · Churchill Medical Systems, Inc. · initiated Jan 21, 2013

Class II Terminated All recalls posted in March 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Churchill Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The label on the device states that the item number is CLGY-2210 when the item number on the device label should read CLGY-2010. This is a 20Ga Huber needle set; however, the product label states the pouch contains a 22Ga set.

FDA root cause category: Under Investigation by firm

What the company is doing

Vycon sent an Urgent Medical Device Recall letter dated January 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed check their stock for the affected product and immediately cease use and distribution and quarantine all affected product immediately. Count and document their affected inventory on the Recall Acknowledgement and Inventory Return Form. Customers were instructed to fax or email the form to Vygon including customer contact information. All affected product would be credited or replaced at no cost to the customer. Customers with questions were instructed to call 1-800-473-5414, or by e-mail at [email protected]. For questions regarding this recall call 603-743-5988.

Recall details

Recall number
Z-0914-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 21, 2013
Posted by the FDA
Mar 6, 2013
Recalling firm
Churchill Medical Systems, Inc.
Quantity
2340
Product code
FPA
FDA submissions
K013871
Distribution
Nationwide Distribution including MO, IL, FL, PA, VA, MO, NY, NE, NH, and OH.

Product description

Lifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 Huber type needle set used for the infusion of medical fluids into implantable ports.

Code information (models, lots, serial numbers)

Lot #1210099

Other devices under the same FDA submission

3 devices cleared under K013871 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.