Genesis Straight Esthetic 5.0 mm 3. mm Cuff Abutment recallZ-0924-2013 · Keystone Dental Inc · initiated Jul 5, 2011
Class II Terminated All recalls posted in March 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Keystone Dental Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mislabeled: Genesis Straight Esthetic 4.0mm Contour ,1mm cuff labeled as Genesis Straight Esthetic 5.0 mm, 3mm cuff abutment
FDA root cause category: Mixed-up of materials/components
What the company is doing
Keystone issued Market Withdrawal notice via fax on July 5, 2011, to affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete and return the fax-back form to facilitate whether they hae any affected product and indicate whether they would like to receive a replacement of a full refund. Customers ewre asked to contact Keyston'es Customer Relations Department to coordinate shipment of the returned product. For questions regarding this recall call 781-328-3390.
Recall details
- Recall number
- Z-0924-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 5, 2011
- Posted by the FDA
- Mar 8, 2013
- Recalling firm
- Keystone Dental Inc
- Quantity
- 22 units
- Product code
- NHA
- FDA submissions
- K101545
- Distribution
- Nationwide Distribution including CA, IN, MA, MD, MN, MI, , NH, NY, NC, PA, and WI.
Product description
Genesis Straight Esthetic 5.0 mm 3. mm Cuff Abutment Part Number: G71205 Genesis implant for single stage or two stage surgical procedures all types of of bone in partially or fully edentulous mandibles and maxillae.
Code information (models, lots, serial numbers)
Lot Number: 2385 Exp Date: 09/2013
Other devices under the same FDA submission
20 devices cleared under K101545 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Genesis Implant System G21138Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM308 reports
- Genesis Implant System G21145Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM198 reports
- Genesis Implant System G21139Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM197 reports
- Genesis Implant System G21144Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM190 reports
- Genesis Implant System G21137Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM186 reports
- Genesis Implant System G21132Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM163 reports
- Genesis Implant System G21131Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM149 reports
- Genesis Implant System G21140Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM138 reports
- Genesis Implant System G21133Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM118 reports
- Genesis Implant System G21130Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM116 reports
- Genesis Implant System G21146Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM75 reports
- Genesis Implant System G21151Keystone Dental, Inc. · IMPLANT, ENDOSSEOUS, ROOT-FORM53 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 64477).
- Z-0923-2013 · TerminatedGenesis Straight Esthetic 4.0 mm 1mm Cuff Abutment Part Number: G71200 Genesis implant for single stage or two stage surgical procedures all types of of bone in partially or fully…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page