Skip to content

ACCU-CHEK Inform II Blood Glucose Monitoring System bar… recallZ-1113-2013 · Roche Diagnostics Operations, Inc. · initiated Mar 6, 2013

Class II Terminated All recalls posted in April 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

FDA root cause category: Device Design

What the company is doing

Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.

Recall details

Recall number
Z-1113-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 6, 2013
Posted by the FDA
Apr 15, 2013
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
Both devices 4467
Product code
NBW
FDA submissions
K121679
Distribution
Nationwide Distribution.

Product description

ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Code information (models, lots, serial numbers)

part number 05060311001, 05060303001

Other devices under the same FDA submission

4 devices cleared under K121679 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 64460).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page