ORTHOCORD One Violet and One Blue Braided Composite Suture… recallZ-1174-2013 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Mar 22, 2013
Class II Terminated All recalls posted in April 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
FDA root cause category: Labeling Change Control
What the company is doing
On March 22, 2013, firm notified consignees of recall via letter. Consignees were informed of the following: The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six weeks after implantation. As Blue and Violet ORTHOCORD free strand sutures share the same IFU, IFU-108147 will be revised. Consignees were requested to confirm the receipt of the letter by March 29, 2013.
Recall details
- Recall number
- Z-1174-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 22, 2013
- Posted by the FDA
- Apr 25, 2013
- Recalling firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Quantity
- 206540
- Product code
- NEW
- FDA submissions
- K043298
- Distribution
- Nationwide Distribution.
Product description
ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Code information (models, lots, serial numbers)
ALL from Jan 1 2010 - Corrected IFU implementation date
Other devices under the same FDA submission
3 devices cleared under K043298 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ORTHOCORD 223114Medos International Sàrl · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE15 reports
- ORTHOCORD 223113Medos International Sàrl · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE5 reports
- ORTHOCORD 223111Medos International Sàrl · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 64706).
- Z-1172-2013 · TerminatedORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Z-1173-2013 · TerminatedORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use in general soft tissue approximation, and/or ligation, including orthoped…
- Z-1175-2013 · TerminatedORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. For use in general soft tissue approximation, and/or ligation, including…
- Z-1176-2013 · TerminatedORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 223116. For use in general soft tissue approximation, and/or ligation,…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page