Suturable DuraGen Dural Regenerative Matrix recallZ-1283-2013 · Integra LifeSciences Corporation · initiated Mar 28, 2013
Class II Terminated All recalls posted in May 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
FDA root cause category: Process control
What the company is doing
The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: [email protected] or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol." For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723.
Recall details
- Recall number
- Z-1283-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 28, 2013
- Posted by the FDA
- May 10, 2013
- Recalling firm
- Integra LifeSciences Corporation
- Quantity
- 4, 799 boxes
- Product code
- GXQ
- FDA submissions
- K043427
- Distribution
- Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.
Product description
Suturable DuraGen Dural Regenerative Matrix Suturable DuraGen Dural Regenerative Matrix is an absorbable implant for repair of dural defects. Suturable DuraGen is an easy to handle, soft, white, pliable, nonfriable, porous collagen matrix with a mechanically strengthened collagen component. Suturable DuraGen is supplied sterile, non-pyrogenic, for single use in double peel packages in a variety of sizes. Suturable DuraGen may be applied using either on-lay or suturing techniques depending on clinical need and surgeon preference.
Code information (models, lots, serial numbers)
1125682, 1125681 1125817, 1125983, 1126018 1126019, 1130641 1130642, 1110848, 1110849 1111883, 1111354, 1111884 1111936, 1112151 1110256, 1111355
Other devices under the same FDA submission
3 devices cleared under K043427 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DuraGen® Suturable DURS2291Integra Lifesciences Corporation · DURA SUBSTITUTE5 reports
- DuraGen® Suturable DURS3391Integra Lifesciences Corporation · DURA SUBSTITUTE3 reports
- DuraGen® Suturable DURS1391Integra Lifesciences Corporation · DURA SUBSTITUTE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 64778).
- Z-1282-2013 · TerminatedDural Graft Matrix DuraGen Dural Graft Matrix is an absorbable implant for repair of dural defects. DuraGen matrix is an easy to handle, soft, white, pliable, nonfriable, porous …
- Z-1284-2013 · TerminatedDuraGen Plus Dural Regeneration Matrix DuraGen Plus Adhesion Barrier Matrix is an absorbable implant for repair and restoration of dural defects and as an adhesion barrier for the…
- Z-1286-2013 · TerminatedNeuraWrap Nerve Protector NeuraWrap Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
- Z-1287-2013 · TerminatedNeuraGen Nerve Guide NeuraGen Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page