AMS 800 Urinary Control System recallZ-1280-2013 · American Medical Systems, Inc. · initiated Mar 1, 2013

Class II Terminated All recalls posted in May 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

American Medical Systems, Inc. is initiating a recall on two components of the AMS 800 Urinary Control System, Control Pump & Pressure-Regulating Balloon. These products were released despite not meeting all final product release criteria.

FDA root cause category: Release of Material/Component prior to receiving test results

What the company is doing

American Medical Systems sent two written communications distributed within the United States. The first is a letter that is being delivered to customers who purchased the affected lots using their ship to address from the original sales order. The second is a letter being delivered to implanting physician customers who are known by AMS to be regular users of the devices. Similar letters will be sent to international customers, including both end users and distributors (where applicable). American Medical Systems sent a "Urgent Medical Device Recall" letter dated February 28, 2013, to all affected customers. The letters identified the product, the problem, and the action needed to be taken by the customer. Customers were instructed to return the recalled product and to complete the Product Recall Acknowledgement Form. The letters addressed to "Physicians" advised consignees against prophylactic removal of implanted control pumps and pressure-regulating balloons contained within the scope of the product recall. For further questions please call AMS Customer Service at 1 (800) 328-3881

Recall details

Recall number
Z-1280-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 1, 2013
Posted by the FDA
May 10, 2013
Recalling firm
American Medical Systems, Inc.
Quantity
77
Product code
EZY
FDA submissions
P000053
Distribution
Worldwide Distribution -- Nationwide Distribution including the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY,NC, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VT, VA, WA and WI., and the countries of AUSTRIA, AUSTRALIA, CANADA, BELGIUM, SWITZERLAND, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, SPAIN, FINLAND, FRANCE, GREAT BRITAIN, IRELAND, ITALY, LEBANON, NETHERLANDS, NORWAY, SWEDEN, TURKEY, ARGENTINA, BRAZIL, IVORY COST, COLOMBIA, ECUADOR, INDIA, JAPAN, MEXICO, NEW ZEALAND, PANAMA and VENEZUELA.

Product description

AMS 800 Urinary Control System. The AMS 800 Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) following prostate surgery. The AMS 800 Urinary Control System is an implantable, fluid filled, solid silicone elastomer device used to treat stress urinary incontinence. It is designed to restore the natural process of urinary control. The device simulates normal sphincter function by opening and closing the urethra at the control of the patient. The AMS 800 system consists of three inter connected components: a cuff, a pump, and a pressure-regulating balloon. The three components are connected with kink-resistant tubing. The AMS 800 can be implanted at either the bulbous urethra or the bladder neck.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model # Lot #  72400024 771029003 72400024 794405011 72400024 744211005 72400024 794405007 72400024 744211008 72400024 771029001 72400024 794405006 72400024 771029005 72400024 744211009 72400024 771029016 72400023 741419006 72400024 794405001 72400024 794405002 72400024 794405004 72400024 794405005 72400024 794405015 72400024 794405016 72400024 744211001 72400024 744211004 72400024 800345007 72400024 800345014 72400024 771029012 72400024 794405009 72400024 794405012 72400024 794405013 72400024 794405014 72400023 741419004 72400023 741419005 72400023 741419007 72400023 741419009 72400023 741419010 72400024 744211002 72400024 744211003 72400024 744211006 72400024 744211007 72400024 744211010 72400024 744211011 72400024 771029002 72400024 771029004 72400024 771029006 72400024 771029007 72400024 771029008 72400024 771029009 72400024 771029010 72400024 771029011 72400025 803589003 72400024 771029014 72400024 771029015 72400025 803589005 72400024 794405003 72400024 771029013 72400024 794405008 72400024 800345008 72400024 794405010 72400024 800345001 72400024 800345003 72400024 800345004 72400024 800345005 72400024 800345006 72400024 800345009 72400024 800345010 72400024 800345012 72400024 800345015 72400024 800345002

Other devices under the same FDA submission

31 devices cleared under P000053 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 64875).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page