epoc BGEM Test Card (Catalogue number CT-1004-00-00) recallZ-1971-2013 · Epocal · initiated Mar 13, 2013

Class II Terminated All recalls posted in August 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Epocal or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On March 13th 2013, Epocal placed all CT-1004-00-00 test card on shipment hold. Affected product in Epocal control has been segregated and quarantined. Beginning on March 26 2013, Epocal Inc. plans to begin distribution of the Recall Notification Letter to the 4 direct consignees (customers/distributors) via certified mail. Included with the Recall Notification Letter is a faxback verification form. Notification will include all distribution channels to the end-user level. Direct consignees are instructed to discontinue use and discard all stock on hand, complete the provided faxback verification form to document disposal and contact their consignees to advise them of the notification and disposal of affected product.

Recall details

Recall number
Z-1971-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 13, 2013
Posted by the FDA
Aug 14, 2013
Recalling firm
Epocal
Quantity
828,750 Test Card
Product code
CGA
FDA submissions
K090109, K093297
Distribution
Worldwide Distribution-USA (nationwide) including the District of Columbia, US Virgin Islands and Puerto Rico, and the states of CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MS, NC, ND, NE, NH, NJ, NM, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VE, WA, WI, and WV and the coutries of India, Argentina, Australia, Azerbaijan, Belarus, Canada, China, Colombia, Egypt, Germany, Guatemala, Israel, Japan, Malaysia, Mexico, Nepal, Netherlands, Peru, Philippines, Russia Fed., Singapore, South Africa, South korea, Thailand, and Vietnam.

Product description

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

LOT #'s EXP. DATES 07-12283-00 03/26/2013 07-12284-00 03/27/2013 07-12285-00 03/28/2013 07-12289-00 04/01/2013 07-12290-00 04/02/2013 07-12292-00 04/04/2013 07-12293-00 04/05/2013 07-12296-00 04/08/2013 07-12297-00 04/09/2013 07-12298-00 04/10/2013 07-12300-00 04/12/2013 07-12303-00 04/15/2013 07-12305-00 04/17/2013 07-12306-00 04/18/2013 07-12307-00 04/19/2013 07-12310-00 04/22/2013 07-12312-00 04/24/2013 07-12313-00 04/25/2013 07-12314-00 04/26/2013 07-12317-00 04/29/2013 07-12320-00 05/02/2013 07-12321-00 05/03/2013 07-12324-00 05/06/2013 07-12325-00 05/07/2013 07-12327-00 05/09/2013 07-12331-00 05/13/2013 07-12333-00 05/15/2013 07-12335-00 05/17/2013 07-12338-00 05/20/2013 07-12340-00 05/22/2013 07-12341-00 05/23/2013 07-12345-00 05/27/2013 07-12346-00 05/28/2013 07-12347-00 05/29/2013 07-12349-00 05/31/2013 07-12352-00 06/03/2013 07-12354-00 06/05/2013 07-12355-00 06/06/2013 07-12362-00 06/13/2013 07-12363-00 06/14/2013 07-13002-00 06/19/2013 07-13003-00 06/20/2013 07-13004-00 06/21/2013 07-13007-00 06/24/2013 07-13008-00 06/25/2013 07-13009-00 06/26/2013 07-13010-00 06/27/2013 07-13014-00 07/01/2013 07-13015-00 07/02/2013 07-13017-00 07/04/2013

Other devices under the same FDA submission

1 device cleared under K090109, K093297 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.