Skip to content

ADVIA Centaur Systems TSH3 Ultra ReadyPack ((REF 06491072 -… recallZ-1881-2013 · Siemens Healthcare Diagnostics, Inc · initiated May 10, 2013

Class II Terminated All recalls posted in August 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A rare variant of TSH, identified in a small cluster of patients, is not detected.

FDA root cause category: Device Design

What the company is doing

On May 10, 2013, an Important Safety Information notice was sent to all affected Siemens Healthcare Diagnostic customers in the United States. On May 10, 2013, an Urgent Field Safety Notice was sent to all customers outside of the United States. These notices inform the customer that there is a rare variant of TSH, identified in a small cluster of patients that is not detected by some Siemens assays. This notice informs customers that they should continue to use the affected TSH assays as the observed rate of occurrence during a 30-month time period was 0.6 X 10 -7 and there are strong mitigating factors. These mitigating factors include co-interpretation of TSH results in conjunction with thyroxine and TS, patient history and clinical signs and symptoms. Although Siemens believes a look back is not required as the expected frequency of the variant is extremely rare, these notices state that this decision should be made by the medical director.

Recall details

Recall number
Z-1881-2013
Recall class
Class II
Status
Terminated
Initiated by the company
May 10, 2013
Posted by the FDA
Aug 6, 2013
Recalling firm
Siemens Healthcare Diagnostics, Inc
Quantity
172913
Product code
JLW
FDA submissions
K083844
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.

Product description

ADVIA Centaur Systems TSH3 Ultra ReadyPack ((REF 06491072 - 100 Tests, REF 06491080 - 500 Tests, REF 04862625 - 2500T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).

Code information (models, lots, serial numbers)

Per recall strategy.

Other devices under the same FDA submission

2 devices cleared under K083844 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page