Dimension Vista 500 and Dimension Vista 1500 System… recallZ-2088-2013 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 19, 2013

Class II Terminated All recalls posted in August 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens is conducting a field correction for Dimension Vista 500 and Dimension Vista 1500 systems using software versions 3.5.1 and 3.6. Under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a QC or calibrator product from a Vista vial. If two samples are dispensed into the same aliquot

FDA root cause category: Software Design Change

What the company is doing

The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated June 2013 to its consignees/customers. The letter describes the product, problem and actions to be taken. The customers were instructed to take the following actions: 1) In the event that an instrument Reset is required for any reason, review the Current Activities Screen to determine if there are any pending QC or Calibration tests. 2) If there are no pending QC of Calibration tests, Rest the instrument or troubleshoot the instrument per the Operators Guide. 3) If there are pending QC or Calibration tests, restarting the software is necessary before resuming patient processing. Go to Operation -> Shutdown, confirm Yes, then on the blue console menu screen select start Vista. Consignees/customers were also asked to complete the Field Effectiveness Check form attached to the letter via fax to Siemens Healthcare Diagnostics at (302) 621-8467, and retain the letter with laboratory records, and forward the letter to those who may have received the product. If you have any questions, contact your Siemens Customer Care Center at (800-441-9250)-Technical Solutions or your local Siemens Technical Support representative.

Recall details

Recall number
Z-2088-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 19, 2013
Posted by the FDA
Aug 26, 2013
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1937
Product code
JJE
FDA submissions
K051087
Distribution
Worldwide distribution: US (nationwide) including PR and countries of: Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudia Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, and United Kingdom.

Product description

Dimension Vista 500 and Dimension Vista 1500 System Aliquot Well Double Dispense with software versions 3.5.1 and 3.6. In vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric, and integrated ion-selective multisensory technology for clinical use.

Code information (models, lots, serial numbers)

Dimension Vista 500 Siemens material numbers: 10284473 (domestic), and 10488224 (international), Dimension Vista 1500 Siemens material numbers: 10444801 (domestic) and 10444802 (international).

Other devices under the same FDA submission

4 devices cleared under K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.