Synthes Large External Fixation, MR Conditional, Tube to… recallZ-1816-2013 · Synthes USA HQ, Inc. · initiated Oct 12, 2012
Class II Open, Classified All recalls posted in July 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
FDA root cause category: Packaging
What the company is doing
Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329. If you have any further questions please call 610-719-5450.
Recall details
- Recall number
- Z-1816-2013
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 12, 2012
- Posted by the FDA
- Jul 29, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 8
- Product code
- KTT
- FDA submissions
- K082650
- Distribution
- US Distribution including the states of CA, MN, NC and TN.
Product description
Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
Code information (models, lots, serial numbers)
Part Number 390.007, Lot Number 6800599
Other devices under the same FDA submission
32 devices cleared under K082650 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 390.005Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT21 reports
- NA 390.008Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE21 reports
- NA 294.56Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 390.003Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 394.85Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 394.86Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT6 reports
- NA 294.786Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT5 reports
- NA 294.53Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT4 reports
- NA 294.57Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE4 reports
- NA 294.785Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT4 reports
- NA 294.52Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE3 reports
- NA 294.56SHASynthes GmbH · Screw, fixation, bone3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page