ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No.… recallZ-2171-2013 · Siemens Healthcare Diagnostics, Inc · initiated Jul 25, 2013
Class II Terminated All recalls posted in September 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Calibration failures and/or significant negative shifts in quality control (QC) and patient results.
FDA root cause category: Material/Component Contamination
What the company is doing
Siemens sent an Urgent Medical Device Correction letter on July 26, 2013, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. an Urgent Field Safety Notice was sent to all domestic customers via Federal Express. On July 25, 2013 this notice was sent to all OUS customers via hard copy, email, and/or fax as determined by each country's local regulations and procedures. These notices inform the customer that a portion of Folate ReadyPacks in kit lots ending in 218 are exhibiting calibration failures and/or significant negative shifts in quality control (QC) and patient results. If customers have an alternate lot of the ADVIA Centaur or ADVIA Centaur CP Folate assay, they are instructed that this lot should be used for future testing. However if no alternative lot is available, quality control (QC) results should be obtained for each individual ReadyPack as a QC or calibration is an indication of an affected pack. An affected pack must be replaced with a different Folate ReadyPack where a valid calibration and acceptable QC values are obtained. These notices should be reviewed by the facility's laboratory or medical director, though no look back is required. Customers were instructed that Fedex Return Receipts will be used to confirm that the customers received the UMDC. Customers with questions were instructed to contact their local Technical Support Provider or Distributor. For questions regarding this recall call 508-668-5000.
Recall details
- Recall number
- Z-2171-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 25, 2013
- Posted by the FDA
- Sep 10, 2013
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 488
- Product code
- CGN
- FDA submissions
- K010050
- Distribution
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada and Mexico.
Product description
ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
Code information (models, lots, serial numbers)
22298218
Other devices under the same FDA submission
3 devices cleared under K010050 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA Centaur® FOL 10310308Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall3 reports
- ADVIA Centaur® FOL 10325366Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall1 report
- ADVIA Centaur® FOL 10331250Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 65806).
- Z-2169-2013 · TerminatedADVIA Centaur Folate (100 Tests), SMN 10310308, Cat No. 06367974 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-2170-2013 · TerminatedADVIA Centaur Folate (500 Tests) SMN 10325366, Cat No. 06891541 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-2172-2013 · TerminatedADVIA Centaur Folate (2500 tests), REF SMN 10340209, Cat No. 09132781 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page