COBAS INTEGRA 800 Analyzer with software 9864.C2 recallZ-0168-2014 · Roche Diagnostics Operations, Inc. · initiated Sep 16, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof
FDA root cause category: Software Design Change
What the company is doing
Roche sent an Urgent Medical Device Correction letter dated September 16, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Roche Diagnostics recommends tracking all ISE counters and manually requesting the corresponding Service Actions listed in the notification be monitored until updated software becomes available during Q4 2013. If the affected product was moved to another site, the notification is to be forwarded. consignees were asked to complete the attached faxback form and fax it to 1-877-266-0997 and file the Urgent Medical Device Correction (UMDC) for future reference. Question should be directed to Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 For questions regarding thisr ecall call 317-576-3911.
Recall details
- Recall number
- Z-0168-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 16, 2013
- Posted by the FDA
- Nov 5, 2013
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 527 units
- Product code
- JJE
- FDA submissions
- K951595
- Distribution
- Nationwide Distribution and Puerto Rico
Product description
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.
Code information (models, lots, serial numbers)
software version 9864.C2
Other devices under the same FDA submission
11 devices cleared under K951595 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE17 reports
- Creatinine Jaff Gen2 6407137190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE13 reports
- Total Protein Gen2 3183734190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN11 reports
- Phosphate (Inorganic) ver.2 03183793122Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus11 reports
- Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 20764949322Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot6 reports
- Ethanol Gen2 3183777190Roche Diagnostics GmbH · ALCOHOL DEHYDROGENASE, SPECIFIC REAGENT FOR ETHANOL ENZYME METHOD5 reports
- Total Protein Gen2 5171385190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN5 reports
- Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 5850819190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot4 reports
- Aspartate Aminotransferase acc to IFCC with pyridoxal phosphate activation 4467493190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot3 reports
- Ethanol Gen.2 08057630190Roche Diagnostics GmbH · ALCOHOL DEHYDROGENASE, SPECIFIC REAGENT FOR ETHANOL ENZYME METHOD2 reports
- Ethanol Gen.2 08445672190Roche Diagnostics GmbH · ALCOHOL DEHYDROGENASE, SPECIFIC REAGENT FOR ETHANOL ENZYME METHOD1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page