Nova StatStrip/StatSensor Hospital Blood Glucose Meter recallZ-0650-2014 · Nova Biomedical Corporation · initiated Oct 10, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nova Biomedical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

FDA root cause category: Component design/selection

What the company is doing

On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.

Recall details

Recall number
Z-0650-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 10, 2013
Posted by the FDA
Jan 9, 2014
Recalling firm
Nova Biomedical Corporation
Quantity
est. 200,000
Product code
CGA
FDA submissions
K060345, K063821, K070068
Distribution
Nationwide Distribution.

Product description

Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Code information (models, lots, serial numbers)

All serial numbers

Other devices under the same FDA submission

3 devices cleared under K060345, K063821, K070068 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.