Trabecular Metal" Reverse Shoulder System recallZ-1079-2014 · Zimmer, Inc. · initiated Oct 17, 2013
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.
FDA root cause category: Device Design
What the company is doing
In November and December 2012 Sales staff were provided updated surgical technique and associated video to share with current customers, surgeons, and risk managers. October 17, 2013 Distributors ( via E-mail),and Risk Managers and Surgeons (via traceable courier) were sent written Urgent Recall notifications of these labeling changes. Current customers are identified through an invoice search and distribution history for the affected devices for the past 12 months, since 9/1/12.
Recall details
- Recall number
- Z-1079-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 17, 2013
- Posted by the FDA
- Feb 24, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 73,002 units
- Product code
- KWT
- FDA submissions
- K052906
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.
Product description
Trabecular Metal" Reverse Shoulder System
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 60568467, 60568466, 60564623, 60564639, 60564628, 60564641, 60564638, 60564636, 60568470, 60564642, 60564635, 60564625, 60564621, and 60564633.
Other devices under the same FDA submission
34 devices cleared under K052906 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trabecular Metal® 00-4349-036-11Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED17 reports
- Trabecular Metal® 00-4349-012-13Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED14 reports
- Trabecular Metal® 00-4349-036-03Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED9 reports
- Trabecular Metal® 00-4349-014-13Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED8 reports
- Trabecular Metal® 00-4349-036-00Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED7 reports
- Trabecular Metal® 00-4349-066-06Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED7 reports
- Trabecular Metal® 00-4349-036-06Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED6 reports
- Trabecular Metal® 00-4349-040-11Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED6 reports
- Trabecular Metal® 00-4309-029-00Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED5 reports
- Trabecular Metal® 00-4349-065-06Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED5 reports
- Trabecular Metal™ 00-4309-041-00Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED4 reports
- Trabecular Metal® 00-4349-039-12Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page