Medtronic RestoreUltra, model 37712, and RestoreSensor… recallZ-0278-2014 · Medtronic Neuromodulation · initiated Sep 30, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
FDA root cause category: Software design
What the company is doing
In September 2013 Medtronic issued a medical device correction letter titled "Loss of Stimulation and Over Stimulation" to customers associated with the affected Medtronic neurostimulators. This letter informed customers of the potential for these issues to occur and provided recommendations to significantly reduce the likelihood of the issues. Medtronic field representatives will be updating the model 8870 software application card to version AAR/01, which is contained in the physician programmer. A Physician Reply Form was asked to be signed and returned by fax or mail for acknowledging the receipt of the Urgent Medical Device Correction letter. When the Medtronic field rep updates the software application card of the physicains' programmer, they will complete a Rep Reply For: 8870 Card Update to AAR/01.
Recall details
- Recall number
- Z-0278-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 30, 2013
- Posted by the FDA
- Nov 14, 2013
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 64,163
- Product code
- LGW
- FDA submissions
- P840001
- Distribution
- Worldwide distribution: US (Nationwide) including Puerto Rico; and countries of: Algeria, Aruba, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malta, Martinique, Mexico, Morocco, Netherlands, Netherlands Antilles, New Caledonia, Norway, Oman, Pakistan, Panama, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.
Product description
Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
Code information (models, lots, serial numbers)
all serial numbers
Other devices under the same FDA submission
116 devices cleared under P840001 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- RestoreSensor 37714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,116 reports
- PrimeAdvanced® SureScan® 97702Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death2,263 reports
- RestoreUltra® SureScan® 97712Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,887 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1,339 reports
- Vanta™ AdaptiveStim™ 977006Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief3 recalls1,151 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy1,125 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy635 reports
- RestoreAdvanced® SureScan® 97713Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF632 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief513 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 66506).
- Z-0279-2014 · TerminatedMedtronic Activa PC, model 37601, and Activa RC, model 37612, Activa SC, models 37602 and 37603. Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page