CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle" recallZ-0099-2015 · Intra Pump Infusion Systems · initiated Oct 18, 2013

Class II Terminated All recalls posted in October 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intra Pump Infusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product has the potential to have a clear debris on the tip of the needle.

FDA root cause category: Device Design

What the company is doing

Cane S.p.A. Medical Technology sent an Urgent: Correction and Removal Action Required letter dated October 21, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: a. Stop using the 19G x 11/2" 1.10 x 40mm steelneedle included in the CRONO Syringe blister pack. B. Safely discard the 19G x 11/2" 1.10 x 40mm steel needle into a Sharps container. c. Use a TRANSFER DEVICE (e.g.,Needleless adapter with female-to-female luer connector) or alternative device to transfer medications into the CRONO syringe. d.For intravenous infusions,we are recommending that a filter also be used during infusions. e. Complete the required 'Acknowledgment and Receipt Form'and fax back to our Distributor;Intra Pump Infusion Systems at 630-845-2768. Contact Information for our US Distributor: INTRA PUMP INFUSION SYSTEMS " Monday through Friday 8:00am to 5:00pm CST,TollFree:1-866- 211-7867 " Direct:630-845-7867 " Fax:630-845-2768 www.lntrapump.com

Recall details

Recall number
Z-0099-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 18, 2013
Posted by the FDA
Oct 21, 2014
Recalling firm
Intra Pump Infusion Systems
Quantity
25,440 pieces
Product code
FRN
FDA submissions
K041414, K042085, K052217, K052218, K052219
Distribution
Worldwide Distribution: US (nationwide) in states of CA, PA, MA, IL, OH and NY; and Internationally to Canada.

Product description

CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"

Code information (models, lots, serial numbers)

Product Code CRN20, lot numbers 14N02, 15E01, 15A02

Other devices under the same FDA submission

1 device cleared under K041414, K042085, K052217, K052218, K052219 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.