Skip to content

Crono Super PIDRecalls & reported problems

Cane' SPA · The CRONO Super PID is an ambulatory syringe infusion pump intended for the controlled, subcutaneous administration of liquids. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08050616612021), received between May 25, 2021 and Aug 8, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Inaccurate Flow Rate” (1 report).

1 other recall cover devices cleared under the same FDA submission (K052217); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Flow Rate1
  • Infusion or Flow Problem1

Patient problems

  • Abdominal Cramps1
  • Anemia1
  • Headache1
  • Nausea1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2024Infusion or Flow Problem; Infusion or Flow Problem

    WE WERE MADE AWARE BY OUR DISTRIBUTOR OF AN EVENT REPORTED ON MAUDE DATABASE (REPORT NUMBER MW5153640) OF WHICH WE HAD NOT BEEN MADE PREVIOUSLY AWARE. EVENT DESCRIPTION: "SPOKE WITH MEMBER'S NURSE, SHE DID INFUSION FOR DEFEROXAMINE, AND IT RAN OVER 8 HOURS JUST FINE PREVIOUS DAY. TODAY MORNING, SHE STARTED THE INFUSION, AND PUMP WAS INFUSING TOO FAST, IN ABOUT 45 MINUTES, ALMOST LESS THAN HALF OF THE MEDICATION WAS ADMINISTERED. SHE STOPPED THE INFUSION. SHE ASKED PHARMACY TO CONTACT HER FOR ASSISTANCE. PHARMACIST REACHED OUT, HOWEVER SHE WAS UNABLE TO REPROGRAM THE PUMP HERSELF, ARRANGED FOR REPLACEMENT PUMP TO BE SENT. MEMBER MISSED HALF OF INFUSION THE DAY OF THE OCCURRENCE OF PUMP FAILURE, HOWEVER NEW ONE WAS SENT FOR NEXT DAY DELIVERY, THEREFORE SHOULD NOT HAVE MISSED ANY FURTHER DOSES. NO ADVERSE EFFECTS REPORTED. PUMP WOULD BE RETURNED ONCE MEMBER RECEIVES THE RETURN BOX. AT THE TIME OF CALL, NO LOT NUMBER OR FURTHER PUMP INFO AVAILABLE. INDICATION: HEMOCHROMATOSIS DUE TO REPEATED RED BLOOD CELL TRANSFUSIONS. REPORTED TO (B)(6) BY: HEALTH PROFESSIONAL".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19939035 ↗
  • InjuryMay 25, 2021Inaccurate Flow Rate; Inaccurate Flow Rate

    WE WERE MADE AWARE BY OUR DISTRIBUTOR OF AN EVENT REPORTED ON MAUDE DATABASE (REPORT NUMBER MW5100144) OF WHICH WE HAD NOT BEEN MADE PREVIOUSLY AWARE. EVENT DESCRIPTION: "SPOKE WITH PATIENT REGARDING ISSUES WITH DEFERRAL AND CRONE PUMP, SHE ADVISED YESTERDAY SHE THOUGHT BATTERY WAS LOW SO SHE TOOK OUT BATTERIES AND PUT THEM BACK IN. SHE NOTICED DURING HER SCHEDULED INFUSION THAT IT STARTED BEEPING AND INFUSION WAS COMPLETED AFTER ONLY 30 MINUTES INSTEAD OF USUAL 8 HOURS. SHE REACHED OUT TO HER NURSE AND WAS ADVISED TO SEEK MEDICAL ATTENTION SINCE SHE REPORTED SIDE EFFECTS OF HEART PALPITATIONS. SHE CALLED ON-CALL MD FOR HEMATOLOGY AND THEY ADVISED SHE SHOULD BE OKAY. SHE WENT TO MD AND SOME BLOOD WORK AND HAD LOW HEMOGLOBIN. SCHEDULED TO RECEIVE A BLOOD TRANSFUSION FOR THIS TOMORROW. SHE SAID EARLIER THAT SHE DID HAVE THE HEART PALPATIONS, HEADACHE, NAUSEA AND STOMACH CRAMPS. SHE ADVISED SHE IS NO LONGER EXPERIENCING THESE SIDE EFFECTS. FOLLOWING UP WITH MD AND NURSING TOMORROW. REPORTED TO (B)(6) BY: PATIENT/CAREGIVER."

    Patient problems: Abdominal Cramps; Anemia; Headache; Nausea

    MAUDE report 11876729 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052217. One submission often covers many models, so check the product and lot details in each record.

  • Z-0099-2015 · Oct 18, 2013 · Terminated
    CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"

Other Pump, Infusion devices

All Pump, Infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page