Genesis(R) Ii, Non-Porous Tibial Base, Size 6, Right, Qty:… recallZ-0131-2014 · Smith & Nephew Inc · initiated Oct 7, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Swapped product. Non-porous tibial base size 5 L was mis-packaged as size 6 R and vice versa.
FDA root cause category: Process control
What the company is doing
Smith & Nephew sent an Urgent - Product Recall 1st Notification letter to all affected sales staff via email and FedEx on October 7, 2013. The hospital was notified by sales rep on October 9, 2013 and the notification letter was sent to the hospital via Fed Ex on October 11, 2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory, discontinue use, quarantine affected products immediately and return to Smith & Nephew. Customers were asked to complete the attached Inventory Return Certification Form and fax to 901-566-7975.
Recall details
- Recall number
- Z-0131-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 7, 2013
- Posted by the FDA
- Nov 1, 2013
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 4 units
- Product code
- JWH
- FDA submissions
- K951987
- Distribution
- USA Nationwide Distribution in the states of: AZ, NC, IA, PA, OH, MD, and TN.
Product description
GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 6, RIGHT, QTY: (1), REF 71420188. Product Usage: orthopaedic
Code information (models, lots, serial numbers)
Batch No. 13GM09505
Other devices under the same FDA submission
137 devices cleared under K951987 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- GENESIS II 71420186Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall38 reports
- GENESIS II 71420166Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall32 reports
- GENESIS II 71420578Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER82 reports
- GENESIS II 71420576Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER72 reports
- GENESIS II 71420566Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER42 reports
- GENESIS II 71420816Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER39 reports
- GENESIS II 71420190Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER35 reports
- GENESIS II 71420182Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER33 reports
- GENESIS II 71420184Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER33 reports
- GENESIS II 71420188Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- GENESIS II 71421032Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- GENESIS II 71926225Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER31 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 66545).
- Z-0130-2014 · TerminatedGENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page