Sigma Lcs High Performance" recallZ-0422-2014 · DePuy Orthopaedics, Inc. · initiated Nov 18, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.
FDA root cause category: Device Design
What the company is doing
DePuy Orthopaedics sent an Urgent Information - Medical Device Correction letter to all affected customers on November 18, 2013. Immediately after initiation on November 18, the company halted initiation with our US distributors and international affiliates because the surgical technique referenced and images from the surgical technique were not correct. No surgeons received the November 18 surgeon letter. The recall was re-initiated on November 20, 2013. The communication to the surgeons who used the affected devices will be delivered by Distributors either by mail or in person and includes the surgeon letter. The device correction includes awareness, reiteration of surgical techniques and instrument care, and visual check of the device as outlined in the Field Notice and Surgeon letter. The devices are not being removed from the market, but the company is investigating potential design changes to the HP M.B.T. Keel Punch Knee Instrument to reduce the potential for the tabs to fracture. All DePuy United States Distributors and DePuy affiliate contacts for Canada, Latin America and International will be notified via email. The US Distributors will identify surgeons who use the affected instruments on Distributor Response Card #1. The US Distributor will notify surgeons either by mail or in person with the written communication. Once all device correction activities are complete, the US Distributor will complete and return Distributor Card #2 to DePuy Orthopaedics, Inc. The US distributors are the US consignees and are responsible for notifying all new HP M.B.T. Keel Punch Knee Instrument surgeon users of this device correction. All territories are required to ensure the HP M.B.T. Keel Punch Knee Instrument Device Correction is promptly initiated. Customers were asked to review the instructions included, along with the attached documents, which contain important information regarding notification of surgeons affected by this Device Correction. Customers
Recall details
- Recall number
- Z-0422-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 18, 2013
- Posted by the FDA
- Nov 27, 2013
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 29,235 units in total / 2008 through November 2013
- Product code
- LXH
- FDA submissions
- K032151, K033272
- Distribution
- Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK
Product description
SIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty.
Code information (models, lots, serial numbers)
Item #950502020
Other devices under the same FDA submission
71 devices cleared under K032151, K033272 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- P.F.C. SIGMA 1581-30-000Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER152 reports
- P.F.C. SIGMA 1581-25-000Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER133 reports
- P.F.C. SIGMA 1581-40-000Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER110 reports
- P.F.C. SIGMA 1581-50-000Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER61 reports
- P.F.C. SIGMA 1581-20-000Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER39 reports
- P.F.C. SIGMA 1581-22-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER23 reports
- P.F.C. SIGMA 1581-23-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER21 reports
- P.F.C. SIGMA 1581-24-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER18 reports
- P.F.C. SIGMA 1581-14-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- P.F.C. SIGMA 1581-14-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- P.F.C. SIGMA 1581-13-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER15 reports
- P.F.C. SIGMA 1581-23-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 66864).
- Z-0412-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED SIZE 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de…
- Z-0413-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or def…
- Z-0414-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin…
- Z-0415-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or …
- Z-0416-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or d…
- Z-0417-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de…
- Z-0418-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definiti…
- Z-0419-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive…
- Z-0420-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive…
- Z-0421-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin…
- Z-0423-2014 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definit…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page