Skip to content

Alaris PC unit model 8000 and 8015 with Alaris PCA module… recallZ-0534-2014 · CareFusion 303, Inc. · initiated Dec 5, 2013

Class I Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

CareFusion is recalling the Alaris PC unit (model 8000 and 8015) because the Alaris PC unit will produce an audible alarm and the Alaris PC unit will display a System Error message.

FDA root cause category: Software design

What the company is doing

CareFusion sent an Urgent Medical Device Recall Notification letter dated December 5, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to contact CareFusion Customer Advocacy at (888) 812-3266, if they experienced a SYSTEM ERROR on the Alaris PC unit. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Customers are instructed to complete and return the enclosed Customer Response Card to expedite the corrective action process. Customers with questions were instructed to contact CareFusion Support Center at 888-562-6018, or for Technical Support call 888-812-3229.

Recall details

Recall number
Z-0534-2014
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Dec 5, 2013
Posted by the FDA
Dec 20, 2013
Recalling firm
CareFusion 303, Inc.
Quantity
4824 total units (4242 units in US)
Product code
FRN
FDA submissions
K041241, K051641
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, and Australia.

Product description

Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.

Code information (models, lots, serial numbers)

not available

Other devices under the same FDA submission

1 device cleared under K041241, K051641 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page