Alaris 8015 (UDI 10885403801518)Recalls & reported problems
Carefusion 303, Inc. · Alaris PC Unit · Class 2
All models of BD Alaris Infusion System (29) · Alaris (25)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is an Alaris 8015 infusion pump unit. Its reports mostly describe corroded or broken parts, failures to read input signals and early battery drain. There are 148,070 reports, including 45 deaths and 233 injuries. Two recalls are open: one covers a connector that may bend and delay infusion or monitoring, the other covers syringe labels listing syringes that were not validated. Reports dropped to 545 in the last 12 months, from 1,572 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 148,079
- Deaths reported
- 45
- Injuries reported
- 234
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 148,079 adverse event reports that list this exact device (DI 10885403801518), received between Oct 23, 2017 and Sep 30, 2026 — classified as 45 deaths, 234 injuries, 147,800 malfunctions.
The most frequently reported specific device problem is “Corroded” (44,537 reports).
2 FDA recall records name this device's identifier, 2 still open.
25 other recalls cover devices cleared under the same FDA submission (K051641); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 15, 2024Class IIOpen, Classified
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Z-3138-2024 · Root cause: Process control · Recall details · FDA record ↗
- Sep 15, 2023Class IOpen, Classified
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Z-0186-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Corroded44,537
- Failure to Read Input Signal42,116
- Break21,932
- Battery Problem8,662
- Premature Discharge of Battery8,525
- Electrical /Electronic Property Problem8,279
- Crack6,158
- Appropriate Device Problem Term/Code Not Available4,690
Patient problems
- No Clinical Signs, Symptoms or Conditions126,088
- No Patient Involvement22,273
- No Consequences Or Impact To Patient436
- Insufficient Information356
- No Information311
- No Known Impact Or Consequence To Patient147
- Low Blood Pressure/ Hypotension91
- Cardiac Arrest16
Recent adverse event reports
The 25 most recent of 148,079 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Premature Discharge of Battery
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942095 ↗MalfunctionSep 29, 2026Break; Contamination; Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931434 ↗MalfunctionSep 24, 2026Electrical /Electronic Property Problem
It was reported that the following issue was observed on the device: "DEFECTIVE PCU KEYPAD ASSEMBLY." Reported Problem Cause Description: "HARDWARE FAILURE." Hence, the work Performed in the device includes: "Found defective PCU keypad assembly and replaced it." There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50882638 ↗MalfunctionSep 24, 2026Communication or Transmission Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Field Service]";"[Channel error]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50880026 ↗MalfunctionSep 21, 2026Communication or Transmission Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Field Service]";"[Channel error]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50837130 ↗MalfunctionSep 18, 2026Corroded
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[service required]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817577 ↗MalfunctionSep 18, 2026Corroded
It was reported that the device may have been affected by water ingress, as the sprinkler system was activated during a fire incident at the hospital, also potential smoke damage. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814858 ↗MalfunctionSep 18, 2026Corroded
It was reported that the device may have been affected by water ingress, as the sprinkler system was activated during a fire incident at the hospital, also potential smoke damage. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50813675 ↗MalfunctionSep 18, 2026Corroded
It was reported that the device may have been affected by water ingress, as the sprinkler system was activated during a fire incident at the hospital, also potential smoke damage. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50813674 ↗MalfunctionSep 18, 2026Corroded
It was reported that the device may have been affected by water ingress, as the sprinkler system was activated during a fire incident at the hospital, also potential smoke damage. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50813664 ↗MalfunctionSep 18, 2026Corroded
It was reported that the device may have been affected by water ingress, as the sprinkler system was activated during a fire incident at the hospital, also potential smoke damage. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50813663 ↗MalfunctionSep 17, 2026Premature Discharge of Battery
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Repair Multi-Device]";"[CM- Repair Multi-Device]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50800951 ↗MalfunctionSep 17, 2026Device Alarm System; Circuit Failure
It was reported that the following issue was observed on the device: "SYSTEM ERROR." Reported Problem Cause Description: "FAULTY LOGIC BOARD." Hence, the work Performed in the device includes: "REPLACED LOGIC BOARD." There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50797780 ↗MalfunctionSep 14, 2026Device Alarm System; Corroded
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[Error code, logic board fault.]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50752253 ↗MalfunctionSep 11, 2026Break
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Repair Multi-Device]";"[Alaris system repairs with a combination of faults.]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50726396 ↗MalfunctionSep 9, 2026Display or Visual Feedback Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[Screen flickering]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50690906 ↗MalfunctionSep 9, 2026Use of Device Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Repair Multi-Device]";"[Alaris system repairs with a combination of faults.]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50688842 ↗MalfunctionSep 9, 2026Corroded
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50686910 ↗MalfunctionSep 8, 2026Circuit Failure
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]"; "[a controller unit SN: (b)(6) with a defective PSU]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50667774 ↗MalfunctionSep 8, 2026Contamination
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Field Service]";"[Equipment fault: Network Comm Error (600.6875)]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50664019 ↗MalfunctionSep 4, 2026Break; Contamination; Corroded; Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50610911 ↗MalfunctionSep 3, 2026Device Alarm System; Corroded
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[faulty 8015 sn 15570100 gives error 800.8000]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589647 ↗MalfunctionSep 3, 2026Battery Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[Fault: battery failure]". There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589630 ↗InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem
It was reported that customer is requesting a log review to rule out any device malfunction following a cardiac arrest event on a patient while potassium was infusing. \\On (b)(6)2026 the patient received potassium supplementation from two sources: 2.4 mEq of potassium as a correction dose infused over 1 hour, and additional potassium delivered through TPN infusion over 24 hours.\\At 17:54, a blood gas analysis showed a potassium level of 2.8 mmol/L. At 18:00, a potassium correction dose of 2.4 mEq was started and infused over 1 hour.\\At 18:40, the patient became agitated with a heart rate of 72 bpm, followed by the development of bradycardia. Cardiopulmonary resuscitation (CPR) was initiated and continued for approximately 10 minutes. Return of spontaneous circulation (ROSC) was achieved after administration of adrenaline, sodium bicarbonate, and noradrenaline.\\At 18:50, the patient's heart rate was 147 bpm. A repeat blood gas analysis showed pH 7.4 and potassium 14.2 mmol/L. Treatment for severe hyperkalemia was initiated with insulin and D10 (10% dextrose), along with Ventolin (salbutamol).\\The patient is currently stable at this time.\\The customer later added the following
Patient problems: Bradycardia; Hyperkalemia
MAUDE report 50584489 ↗MalfunctionSep 2, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50569749 ↗
FDA clearance and classification
Cleared through 510(k) K051641 on Oct 20, 2005 (decision: Substantially Equivalent). Submitted by Cardinal Health, Alaris Products as “Alaris System Pc Unit, Model 8001”. The FDA review took 122 days. FDA record ↗
The FDA regulates product code FRN (pump, infusion) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051641. One submission often covers many models, so check the product and lot details in each record.
- Z-1645-2026 · Feb 11, 2026 · Open, ClassifiedBD Alaris" System with Guardrails" Suite MX with Point of Care Unit
- Z-0185-2024 · Sep 15, 2023 · Open, ClassifiedBD Alaris Syringe Module, REF 8110
- Z-0184-2024 · Sep 15, 2023 · Open, ClassifiedBD Alaris PCU REF 8015
- Z-1361-2020 · Feb 4, 2020 · Open, ClassifiedCareFusion Alaris Syringe Module, Model 8110
- Z-1362-2020 · Feb 4, 2020 · Open, ClassifiedCareFusion Alaris PCA Module, Model 8120
- Z-1360-2020 · Feb 4, 2020 · Open, ClassifiedCareFusion Alaris Pump Module, Model 8100
- Z-2671-2017 · Jun 12, 2017 · TerminatedAlaris PC Unit, Model 8015
- Z-1520-2017 · Nov 1, 2016 · TerminatedAlaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System.
- Z-2064-2016 · May 12, 2016 · TerminatedAlaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central programming, monitoring and power supply component for the Alaris System.
- Z-1606-2016 · Mar 24, 2016 · TerminatedAlaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
- Z-1239-2016 · Feb 12, 2016 · TerminatedAlaris PC Unit, Infusion Pump Model 8000, Part No. TC10005092
- Z-1811-2015 · May 13, 2015 · TerminatedAlaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris
- Z-1311-2015 · Mar 12, 2015 · TerminatedAlaris PC units, Model No. 8015. Infusion pump.
- Z-0534-2014 · Dec 5, 2013 · TerminatedAlaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitor
- Z-1835-2013 · Jun 17, 2013 · TerminatedAlaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part number TC10006584. The Alaris PC Unit is the main user interface unit and power supp
…and 10 more.
Common questions about the Alaris 8015
Has the Alaris 8015 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-3138-2024 (Class II), began on Aug 15, 2024: Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Alaris 8015?
The FDA's MAUDE database lists 148,079 adverse event reports for this device, received between Oct 23, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Corroded” (44,537), “Failure to Read Input Signal” (42,116) and “Break” (21,932); the most common patient problems are “Low Blood Pressure/ Hypotension” (91) and “Cardiac Arrest” (16). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alaris 8015?
45 reports are classified as deaths, 234 as injuries and 147,800 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Alaris 8015 FDA approved?
It was cleared through 510(k) K051641 on Oct 20, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Alaris 8015 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Alaris 8015, and is it still sold?
The FDA's device database lists Carefusion 303, Inc. as the company responsible for it (brand Alaris, model 8015). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,412 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths50,284 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,947 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,545 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,013 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,232 reports
- CADD 21-6400-51Icu Medical, Inc. · Pump, infusion7 deaths7,543 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page