SYSTEM 1E Liquid Chemical Sterilant Processing System… recallZ-0616-2014 · Steris Corporation · initiated Nov 21, 2013

Class II Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.

FDA root cause category: Device Design

What the company is doing

STERIS sent an Urgent Voluntary Field Correction Notice dated November 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that a STERIS Field Service Representative would contact their facility to schedule installation of the new hardware and updated software on their SYSTEM IE Liquid Chemical Sterilant Processing System(s). Customers were instructed that they may continue to operate the SYSTEM IE processor prior to the completion of the update. Customers with questions were instructed to contact STERIS Field Service Dispatch at 1-800-333-8828. For questions regarding this recall call 440-392-7519.

Recall details

Recall number
Z-0616-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 21, 2013
Posted by the FDA
Dec 31, 2013
Recalling firm
Steris Corporation
Quantity
6,028 units
Product code
MLR
FDA submissions
K090036, K101409, K102462, K113520, K131078
Distribution
Nationwide Distribution - USA (nationwide) and the District of Columbia and Puerto Rico. Product was also shipped to the following countries: Afghanistan, Bahrain, Hong Kong, Italy, Saudi Arabia, Turkey & United Kingdom.

Product description

SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Code information (models, lots, serial numbers)

Model #6500 & 7500

Other devices under the same FDA submission

4 devices cleared under K090036, K101409, K102462, K113520, K131078 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.