Skip to content

Siemens Artis Q and Artis Q biplane systems recallZ-1202-2014 · Siemens Medical Solutions USA, Inc · initiated Dec 5, 2013

Class II Terminated All recalls posted in March 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40).

FDA root cause category: Device Design

What the company is doing

Siemens sent an Important Customer Safety Notice letter dated November 22, 2013, to affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to discontinue use until the affected product has been successfully replaced by a Siemen's representative.at no cost.

Recall details

Recall number
Z-1202-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 5, 2013
Posted by the FDA
Mar 27, 2014
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
19
Product code
OWB
FDA submissions
K123529
Distribution
USA Nationwide Distribution in the states of TN, MA, CA, IL, OH, FL, UT, WA, GA, IA, NY, MO, IN, and CO.

Product description

Siemens Artis Q and Artis Q biplane systems Product Usage: Interventional Fluoroscopic X-ray

Code information (models, lots, serial numbers)

model numbers 10848280, 10848281, 10848282, and 10848283.

Other devices under the same FDA submission

7 devices cleared under K123529 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.